Instructions
Chapter 2 Command System  RIGOL 
MSO5000 Programming Guide  2-203   
:TRIGger:IIC:CLEVel 
Syntax 
:TRIGger:IIC:CLEVel <level> 
:TRIGger:IIC:CLEVel? 
Description 
Sets or queries the trigger level of the clock line in I2C trigger. The unit is the same 
as that of the current amplitude. 
Parameter 
Name 
Type 
Range 
Default 
<level> 
Real 
(-5 ×  VerticalScale - OFFSet) to   
(5 ×  VerticalScale - OFFSet) 
0 V 
Remarks 
For VerticalScale, refer to the :CHANnel<n>:SCALe command; for OFFSet, refer to 
the :CHANnel<n>:OFFSet command. 
Return 
Format 
The query returns the trigger level in scientific notation. 
Example 
:TRIGger:IIC:CLEVel 0.16        /*Sets the trigger level to 160 mV.*/ 
:TRIGger:IIC:CLEVel?              /*The query returns 1.600000E-1.*/ 
:TRIGger:IIC:DLEVel 
Syntax 
:TRIGger:IIC:DLEVel <level> 
:TRIGger:IIC:DLEVel? 
Description 
Sets or queries the trigger level of the data line in I2C trigger. The unit is the same 
as that of the current amplitude. 
Parameter 
Name 
Type 
Range 
Default 
<level> 
Real 
(-5 ×  VerticalScale - OFFSet) to   
(5 ×  VerticalScale - OFFSet) 
0 V 
Remarks 
For VerticalScale, refer to the :CHANnel<n>:SCALe command; for OFFSet, refer to 
the :CHANnel<n>:OFFSet command. 
Return 
Format 
The query returns the trigger level in scientific notation. 
Example 
:TRIGger:IIC:DLEVel 0.16        /*Sets the trigger level to 160 mV.*/ 
:TRIGger:IIC:DLEVel?              /*The query returns 1.600000E-1.*/ 