Instructions
Chapter 2 Command System  RIGOL 
MSO5000 Programming Guide  2-3   
:SINGle 
Syntax 
:SINGle 
Description 
Sets the trigger mode of the oscilloscope to "Single". This command functions the same as 
either of the following two operation: press SINGLE on the front panel; or send 
the :TRIGger:SWEep SINGle command. 
Remarks 
  In the single trigger mode, the oscilloscope performs a single trigger when the trigger 
conditions are met and then it stops. 
  When the waveform recording function is enabled or the recorded waveforms are 
played back, this command is invalid. 
  For the single trigger, you can use the :TFORce command to generate one trigger by 
force. 
:TFORce 
Syntax 
:TFORce 
Description 
Generates a trigger signal forcefully. This command is only applicable to the normal and 
single trigger modes (refer to the :TRIGger:SWEep command). This command functions 
the same as the FORCE key in the trigger control area of the front panel. 