Instructions
Chapter 2 Command System  RIGOL 
MSO5000 Programming Guide  2-139   
T
t
Duty Cycle=t/T*100%
Return 
Format 
The query returns the current duty cycle in scientific notation.   
Example 
:PULSe:DCYCle 50        /*Sets the duty cycle of GI pulse to 50%.*/ 
:PULSe:DCYCle?            /*The query returns 5.000000E+1.*/ 
[:SOURce[<n>]]:MOD:TYPE 
Syntax 
[:SOURce[<n>]]:MOD:TYPE <type> 
[:SOURce[<n>]]:MOD:TYPE? 
Description 
Sets or queries the modulation type of the specified function/arbitrary waveform 
generator channel. 
Parameter 
Name 
Type 
Range 
Default 
[<n>] 
Discrete 
{1|2} 
When omitted, by default, the operations are 
carried out on GI. 
<type> 
Discrete 
{AM|FM|FSK} 
AM 
Remarks 
  AM: Amplitude Modulation. The amplitude of the carrier waveform changes with the 
amplitude of the modulating waveform.   
FM: Frequency Modulation. The frequency of the carrier waveform changes with the 
the modulating waveform. 
FSK: Frequency-shift Keying Modulation. Control the carrier frequency changes with 
the digital signal.   
  The Sine, Square, Ramp, Built-in Waveform, or Non-DC signal's arbitrary waveforms 
can be regarded as carrier waveforms. 
  You can select Sine, Square, Triangle, or Noise as the modulating waveforms.   
Return 
Format 
The query returns AM, FM, or FSK. 
Example 
:MOD:TYPE AM        /*Sets the modulation type of GI to AM.*/ 
:MOD:TYPE?            /*The query returns AM.*/ 