RIGOL User’s Guide MSO2000A/DS2000A Series Digital Oscilloscope May 2014 RIGOL Technologies, Inc.
RIGOL Guaranty and Declaration Copyright © 2013 RIGOL Technologies, Inc. All Rights Reserved. Trademark Information RIGOL is a registered trademark of RIGOL Technologies, Inc. Publication Number UGA18105-1110 Notices RIGOL products are covered by P.R.C. and foreign patents, issued and pending. RIGOL reserves the right to modify or change parts of or all the specifications and pricing policies at company’s sole decision.
RIGOL Safety Requirement General Safety Summary Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injury or damage to the instrument and any product connected to it. To prevent potential hazards, please use the instrument only specified by this manual. Use Proper Power Cord. Only the power cord designed for the instrument and authorized for use within the local country could be used. Ground the Instrument.
RIGOL Do Not Insert Anything Into the Holes of Fan. Do not insert anything into the holes of the fan to avoid damaging the instrument. Use Proper Fuse. Please use the specified fuses. Avoid Circuit or Wire Exposure. Do not touch exposed junctions and components when the unit is powered. Do Not Operate With Suspected Failures. If you suspect damage occurs to the instrument, have it inspected by qualified service personnel before further operations.
RIGOL Proper Use of Battery. If a battery is supplied, it must not be exposed to high temperature or in contact with fire. Keep it out of the reach of children. Improper change of battery (note: lithium battery) may cause explosion. Use RIGOL specified battery only. Handling Safety. Please handle with care during transportation to avoid damage to buttons, knob interfaces and other parts on the panels. Safety Terms and Symbols Terms Used in this Manual.
RIGOL Allgemeine Sicherheits Informationen Überprüfen Sie diefolgenden Sicherheitshinweise sorgfältigumPersonenschädenoderSchäden am Gerätundan damit verbundenen weiteren Gerätenzu vermeiden.Zur Vermeidung vonGefahren, nutzen Sie bitte das Gerät nur so, wiein diesem Handbuchangegeben. Um Feuer oder Verletzungen zu vermeiden, verwenden Sie ein ordnungsgemäßes Netzkabel. Verwenden Sie für dieses Gerät nur das für ihr Land zugelassene und genehmigte Netzkabel. Erden des Gerätes.
RIGOL in Betrieb ist. Betreiben Sie das Gerät nicht im Fehlerfall Wenn Sie am Gerät einen Defekt vermuten, sorgen Sie dafür, bevor Sie das Gerät wieder betreiben, dass eine Untersuchung durch qualifiziertes Kundendienstpersonal durchgeführt wird.Jedwede Wartung, Einstellarbeiten oder Austausch von Teilen am Gerät, sowie am Zubehör dürfen nur von RIGOL autorisiertem Personal durchgeführt werden.
RIGOL Sicherheits Begriffe und Symbole Begriffe in diesem Guide. Diese Begriffe können in diesem Handbuch auftauchen: WARNING Die Kennzeichnung WARNING beschreibt Gefahrenquellen die leibliche Schäden oder den Tod von Personen zur Folge haben können. CAUTION Die Kennzeichnung Caution (Vorsicht) beschreibt Gefahrenquellen die Schäden am Gerät hervorrufen können. Begriffe auf dem Produkt.
RIGOL Measurement Category Measurement Category MSO2000A/DS2000A series digital oscilloscopes can make measurements in Measurement Category I. WARNING This oscilloscope can only be used for measurements within its specified measurement categories. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS. Examples are measurements on circuits not derived from MAINS, and specially protected (internal) MAINS derived circuits.
RIGOL Ventilation Requirement This oscilloscope uses fan to force cooling. Please make sure that the air intake and exhaust areas are free from obstructions and have free air. When using the oscilloscope in a bench-top or rack setting, provide at least 10 cm clearance beside, above and behind the instrument for adequate ventilation. WARNING Inadequate ventilation may cause temperature increase which would damage the instrument.
RIGOL Working Environment Temperature Operating: 0℃ to +50℃ Non-operating: -40℃ to +70℃ Humidity 0℃ to +30℃: ≤95% relative humidity +30℃ to +40℃: ≤75% relative humidity +40℃ to +50℃: ≤45% relative humidity WARNING To avoid short circuit inside the instrument or electric shock, please do not operate in humid environment. Altitude Operating: less than 3 km Non-operating: less than 15 km Installation (overvoltage) Category This product is powered by mains conforming to installation (overvoltage) category II.
RIGOL Pollution Degree Degree 2 Pollution Degree Definitions Pollution degree 1: No pollution or only dry, non-conductive pollution occurs. The pollution has no influence. For example: a clean room or air-conditioned office environment. Pollution degree 2: Normally only dry, non-conductive pollution occurs. Occasionally a temporary conductivity caused by condensation may occur. For example: general indoor environment.
RIGOL General Care and Cleaning General Care: Do not store or leave the instrument where it may be exposed to direct sunlight for long periods of time. Cleaning: Clean the instrument regularly according to its operating conditions. To clean the exterior surface: 1. Disconnect the instrument from all power sources. 2. Clean the loose dust on the outside of the instrument with a lint-free cloth (with mild detergent or water). When cleaning the LCD, take care to avoid scarifying it.
RIGOL Environmental Considerations The following symbol indicates that this product complies with the WEEE Directive 2002/96/EC. Product End-of-Life Handling The equipment may contain substances that could be harmful to the environment or human health. In order to avoid the release of such substances into the environment and harm to human health, we encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused or recycled appropriately.
RIGOL MSO2000A/DS2000A Series Overview MSO2000A is a mixed signal digital oscilloscope aimed at embedded system design and test. It allows users to measure both analog and digital signals. MSO2000A/DS2000A series is a multifunctional and high-performance digital oscilloscope designed on the basis of the UltraVision technique developed by RIGOL. It fulfills the design, debugging and test requirements of the mainstream application market.
RIGOL Support USB storage device storage and PictBridge printer Conform to LXI-C class instrument standards; enable quick, economic and efficient creation and reconfiguration of test system Support remote command control User-friendly design 8.
RIGOL Document Overview Main Topics of this Manual: Chapter 1 Quick Start Introduce the preparations before using the oscilloscope and provide a basic introduction of the instrument. Chapter 2 To Set the Vertical System Introduce the vertical system functions of the oscilloscope. Chapter 3 To Set the Horizontal Introduce the horizontal system functions of System the oscilloscope. Chapter 4 To Set the Sample System Introduce the sample system functions of the oscilloscope.
RIGOL Chapter 17 Troubleshooting Chapter 18 Specifications Chapter 19 Appendix remotely. Introduce how to deal with common failures of the oscilloscope. Provide the specifications and general specifications of the oscilloscope. Provide common information such as options and accessories. Format Conventions in this Manual: 1. Key The front panel keys are denoted by the format of “Button Name (Bold) + Text Box”, for example, Utility denotes the “Utility” key. 2.
RIGOL Content Conventions in this Manual: MSO2000A/DS2000A series includes the following models. This manual takes MSO2302A-S for an example to illustrate the functions and performances of MSO2000A/DS2000A series.
Contents RIGOL Contents Guaranty and Declaration .........................................................................I Safety Requirement ................................................................................ II General Safety Summary ........................................................................... II Safety Terms and Symbols ....................................................................... IV Allgemeine Sicherheits Informationen ............................................
RIGOL Contents AUTO ........................................................................................... 1-16 Knob ............................................................................................ 1-17 Navigation Knob ............................................................................ 1-17 MENU ........................................................................................... 1-18 Signal Source ............................................................................
Contents RIGOL High Resolution .............................................................................. 4-4 Sample Mode ........................................................................................ 4-5 Sample Rate ......................................................................................... 4-6 LA Sample Rate ..................................................................................... 4-7 Memory Depth ...................................................................
RIGOL Contents Multiplication ................................................................................... 6-3 Division .......................................................................................... 6-3 FFT ................................................................................................ 6-4 Logic Operation ............................................................................... 6-7 Advanced Operation.................................................................
Contents RIGOL CAN Decoding (Option) .........................................................................8-17 Chapter 9 Reference Waveform ............................................................ 9-1 To Enable REF Function.......................................................................... 9-2 To Set the Color .................................................................................... 9-2 To Select REF Source .........................................................................
RIGOL Contents To Output Ramp ............................................................................ 13-5 To Output Pulse ............................................................................. 13-6 To Output DC ................................................................................ 13-6 To Output Noise ............................................................................ 13-7 To Output Built-in Waveform .................................................................
RIGOL Contents Self-calibration............................................................................. 15-11 Print Setting ................................................................................ 15-12 Power Status ............................................................................... 15-15 Aux Output ................................................................................. 15-16 Option Management.....................................................................
Chapter 1 Quick Start RIGOL Chapter 1 Quick Start This chapter introduces the precautions when using the oscilloscope for the first time, the front/rear panels of the oscilloscope, the user interface and the using method of the built-in help system.
RIGOL Chapter 1 Quick Start General Inspection 1. Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has passed both electrical and mechanical tests. The consigner or carrier shall be liable for the damage to instrument resulting from shipment. RIGOL would not be responsible for free maintenance/rework or replacement of the unit. 2. Inspect the instrument.
RIGOL Chapter 1 Quick Start Appearance and Dimensions 179.6 361.
RIGOL Chapter 1 Quick Start To Prepare the Oscilloscope for Use To Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement of the oscilloscope as well as better operation and observation. Figure 1-3 To Adjust the Supporting Legs To Connect to Power Supply The power requirements of the oscilloscope are 100-240 V, 45-440 Hz.
RIGOL Chapter 1 Quick Start Power-on Inspection When the oscilloscope is energized, pressing the power key at the lower-left corner of the front panel can start the oscilloscope if the power status is currently set to “Default” and the oscilloscope will starts directly if the power status is currently set to “Open”. During the start-up process, the oscilloscope performs a series of self-tests and you can hear the sound of relay switching. After the self-test, the welcome screen is displayed.
RIGOL Chapter 1 Quick Start Figure 1-5 To Connect the Passive Probe Connect the logic probe: 1. Connect the single-wire terminal of the logic probe to the [LOGIC D0-D15] digital channel interface at the front panel of MSO2000A in the correct direction. 2. Connect the signal under test to the other terminal of the logic probe. MSO2000A is provided with an RPL2316 logic probe which provides three connecting methods with the signal under test to fulfill the requirements of different application environment.
RIGOL Chapter 1 Quick Start Function Inspection 1. 2. 3. Press Storage Default to restore the oscilloscope to its factory states. Connect the ground alligator clip of the probe to the “Ground Terminal” as shown in the figure below. Use the probe to connect CH1 of the oscilloscope and the “Probe Compensation Signal Output Terminal”. Probe Compensation Signal Output Terminal Ground Terminal Figure 1-7 Probe Compensation Signal Output Terminal/Ground Terminal 4. 5. Press AUTO.
RIGOL Chapter 1 Quick Start Probe Compensation When the probes are used for the first time, you should compensate the probes to match the input channels of the oscilloscope. Non-compensated or poorly compensated probes may cause measurement inaccuracy or error. The probe compensation procedures are as follows. 1. 2. Perform steps 1, 2, 3 and 4 of “Function Inspection”. Check the displayed waveforms and compare them with the following figures.
RIGOL Chapter 1 Quick Start Front Panel Overview 1 2 13 14 15 3 16 17 18 4 19 5 6 7 20 8 21 9 10 22 11 12 23 Figure 1-10 Front Panel Overview MSO2000A/DS2000A User’s Guide 1-9
RIGOL Chapter 1 Quick Start Table 1-1 Front Panel Descriptions No. Description No.
RIGOL Chapter 1 Quick Start Rear Panel Overview 1 4 5 2 6 7 3 8 Figure 1-11 Rear Panel Overview 1. Handle Pull up the handle vertically for easy carrying of the instrument. When you do not need the handle, press it down. 2. LAN Connect the instrument to the network via this interface for remote control. This oscilloscope conforms to the LXI-C class instrument standards and you can quickly build test system using it with other instruments.
RIGOL Chapter 1 Quick Start (2) Insert a small straight screwdriver into the slot at the power socket and pry out the fuse seat gently. (3) Take out the fuse and replace it with a fuse of the specified specification. Then, reinstall the fuse seat. 5. AC Power Socket AC power input terminal. The power requirement of this oscilloscope is 100-240 V, 45-440 Hz. Use the power cord provided with the accessories to connect the instrument to AC power.
RIGOL Chapter 1 Quick Start Front Panel Function Overview VERTICAL CH1, CH2: analog input channels. The 2 channels are marked by different colors which are also used to mark both the corresponding waveforms on the screen and the channel input connectors. Press any key to open the corresponding channel menu and press again to turn off the channel. MATH: press this key to open the math operation menu under which add, subtract, multiply, divide, FFT, logic and advanced operations are provided.
RIGOL Chapter 1 Quick Start HORIZONTAL MENU: press this key to open the horizontal control menu under which you can turn on or off the delayed sweep function, switch between different time base modes, switch between “Coarse” and “Fine” adjustment of scale as well as modify the horizontal reference setting. HORIZONTAL SCALE: modify the horizontal time base. Turn clockwise to reduce the time base and turn counterclockwise to increase.
RIGOL Chapter 1 Quick Start TRIGGER MODE: press this key to switch the trigger mode to Auto, Normal or Single and the corresponding state backlight of the current trigger mode would be illuminated. TRIGGER LEVEL: modify the trigger level. Turn clockwise to increase the level and turn counterclockwise to reduce. During the modification, the trigger level line would move up or down and the value in the trigger level message box (e.g. ) at the lower-left corner of the screen would change accordingly.
RIGOL Chapter 1 Quick Start CLEAR Press this key to clear all the waveforms on the screen. If the oscilloscope is in the “RUN” state, new waveforms will still be displayed. RUN/STOP Press this key to set the state of the oscilloscope to “RUN” or “STOP”. In the “RUN” state, the key is illuminated in yellow. In the “STOP” state, the key is illuminated in red. SINGLE Press this key to set the trigger mode to “Single” and this key is illuminated in orange.
RIGOL Chapter 1 Quick Start Knob Adjust waveform brightness: In non-menu-operation mode (menu is hidden), turn this knob to adjust the brightness of waveform. The adjustable range is from 0% to 100%. Turn clockwise to increase the brightness and counterclockwise to reduce. Press down this knob to reset the brightness to 50%. You can also press Display WaveIntensity and use the knob to adjust the waveform brightness.
RIGOL Chapter 1 Quick Start MENU Measure: press this key to open the measurement setting menu. You can set the measurement setting, all measure, statistic function etc. Press MENU at the left of the screen to open the measurement menus of 29 waveform parameters. Then, press down the corresponding menu softkey to quickly realize one-key measurement and the measurement result will be displayed at the bottom of the screen.
Chapter 1 Quick Start RIGOL Signal Source Press this key to enter the source setting interface. You can enable or disable the output of the [Source1] or [Source2] connector at the rear panel, set the output signal waveform and parameters, turn on or off the state display of the current signal source. Note: This function is only available for MSO2000A-S and DS2000A-S models oscilloscopes. Record Stop Play/Pause Record Record: press this key to start recording the waveform.
RIGOL Chapter 1 Quick Start Print Press this key to execute the print function or save the content displayed in the screen in the USB storage device in a picture file. If a PictBridge printer is currently connected and the printer is in idle state, pressing this key will execute the print function. If no printer but a USB storage device is currently connected, pressing this key will save the screen content to the USB storage device in “.png” format.
RIGOL Chapter 1 Quick Start User Interface MSO2000A/DS2000A provides 8.0 inch, WVGA (800*480) 160,000 color TFT LCD. What is worth mentioning is that the 14-grid ultra-wide screen enables you to view “longer” waveform. 1 2 12 13 3 14 4 5 15 6 7 16 17 18 Figure 1-12 User Interface 8 9 10 11 19 20 1. Auto Measurement Items Provide 16 horizontal (HORIZONTAL) parameters and 13 vertical (VERTICAL) parameters.
RIGOL Chapter 1 Quick Start level in green (correspond to the color of the channel label). Its edge is displayed in white. The label and waveform of the digital channel currently selected are displayed in red. Note: This function is only applicable to MSO2000A and MSO2000A-S models oscilloscopes. 3. Status Available states include RUN, STOP, T’D (triggered), WAIT and AUTO. 4. Horizontal Time Base Represent the time per grid on the horizontal axis on the screen.
RIGOL Chapter 1 Quick Start 9. Trigger Type Display the currently selected trigger type and trigger condition setting. Different labels are displayed when different trigger types are selected. For example: represents triggering on the rising edge in “Edge” trigger. 10. Trigger Source Display the trigger source currently selected (CH1, CH2, EXT, AC Line or any channel of D0-D15).
RIGOL Chapter 1 Quick Start 14. CH2 Vertical Scale Display the voltage value per grid of CH2 waveform vertically. In addition, the following labels will be displayed according to the current channel setting: channel coupling (e.g. ), input impedance (e.g. ) and bandwidth limit (e.g. ). You can use VIRTICAL SCALE to modify this parameter. 15. Digital Channel Status Area Display the current status of the 16 digital channels (D0 to D15 from right to left).
RIGOL Chapter 1 Quick Start System Time Displayed in “hh:mm (hour:minute)” format. When printing or storing the waveform, the output file will contain this time message. Press Utility System System Time System Time to set through the following format: yyyy-mm-dd hh:mm:ss (year-month-date hour:minute:second) Sound Icon When the sound is enabled, will be displayed. Press Utility Sound to enable or disable the sound.
RIGOL Chapter 1 Quick Start To Use the Security Lock If needed, you can use the security lock (please buy it yourself) to lock the oscilloscope to a fixed location. The method is as follows, align the lock with the lock hole and plug it into the lock hole vertically; turn the key clockwise to lock the oscilloscope and then pull the key out. Security Lock Hole Figure 1-13 To Use the Security Lock Note: Please do not insert other articles into the security lock hole to avoid damaging the instrument.
RIGOL Chapter 1 Quick Start To Use the Built-in Help System The help system of this oscilloscope provides instructions for all the function keys and menu keys at the front panel. Press Help to open the help interface and press again to close the interface. The help interface mainly consists of two parts. The left is “Help Options” and you can use “Button” or “Index” mode to select. The right is “Help Display Area”. Help Options Help Display Area Figure 1-14 Help Interface Button: Default mode.
Chapter 2 To Set the Vertical System RIGOL Chapter 2 To Set the Vertical System The contents of this chapter: To Enable the Analog Channel Channel Coupling Bandwidth Limit Probe Ratio Input Impedance Waveform Invert Vertical Scale Vertical Expansion Amplitude Unit Channel Label Delay Calibration of the Analog Channel MSO2000A/DS2000A User’s Guide 2-1
RIGOL Chapter 2 To Set the Vertical System To Enable the Analog Channel MSO2000A/DS2000A provides 2 analog input channels (CH1 and CH2) and provides independent vertical control system for each channel. As the setting methods of the vertical systems of the two channels are the same, this chapter takes CH1 as an example to illustrate the setting method of the vertical system.
RIGOL Chapter 2 To Set the Vertical System Channel Coupling The undesired signals can be filtered out by setting the coupling mode. For example, the signal under test is a square waveform with DC offset. When the coupling mode is “DC”: the DC and AC components can pass the channel. When the coupling mode is “AC”: the DC components of the signal under test are blocked. When the coupling mode is “GND”: the DC and AC components of the signal under test are both blocked.
RIGOL Chapter 2 To Set the Vertical System Probe Ratio You can set the probe attenuation ratio manually. The probe ratio values available are as shown in the table below. Table 2-1 Probe Attenuation Coefficient Attenuation Coefficient (Amplitude of Signal Under Test: Menu Amplitude of Displayed Waveform) 0.01X 1:100 0.02X 1:50 0.05X 1:20 0.1X 1:10 0.2X 1:5 0.
Chapter 2 To Set the Vertical System RIGOL Input Impedance This oscilloscope provides two input impedance modes (1 MΩ (default) and 50 Ω) to reduce the circuit load caused by the interaction of the oscilloscope and the circuit to be tested. 1MΩ: the input impedance of the oscilloscope is rather high and the current flowing into the oscilloscope from the circuit under test can be ignored. 50Ω: match the oscilloscope with a device with 50 Ω output impedance.
RIGOL Chapter 2 To Set the Vertical System Vertical Scale Vertical scale refers to the voltage value per grid in the vertical direction on the screen and is usually expressed as V/div. Press CH1 Volts/Div to select the desired adjustment mode of the vertical scale, or press VERTICAL SCALE to switch the adjustment mode quickly. Coarse adjustment: set the vertical scale in 1-2-5 step namely 500 μV/div, 1 mV/div, 2 mV/div, 5 mV/div, 10 mV/div…10 V/div (take counterclockwise as an example).
Chapter 2 To Set the Vertical System RIGOL Vertical Expansion When using VERTICAL SCALE to change the vertical scale of the analog channel, you can choose to expand or compress the signal vertically around the center of the screen or the ground point of the signal. Press Utility System VerticalExp to select “Center” or “Ground” and the default is “Ground”. Center: when the vertical scale is modified, the waveform will be expanded or compressed around the center of the screen.
RIGOL Name Input Area Chapter 2 To Set the Vertical System Keyboard Upper/Lower Case Switch Figure 2-2 Label Modification Interface After finishing the input, press OK to finish the modification and the channel label will change to “ ”. To modify or delete the input character, press Name to select the “Name Input Area” and use to select the character to be modified or deleted. Enter the desired character to modify the character or press Delete to delete the character selected.
Chapter 2 To Set the Vertical System RIGOL For MSO2000A/DS2000A, users can set a delay time to calibrate the zero offset of the corresponding channel. Press CH1 Delay-Cal and use to set the desired delay time. The range available is from -200 ns to 200 ns. Pressing down can restore the delay time to 0.00 s. Note: This parameter is related to the instrument model and the current horizontal time base setting.
Chapter 3 To Set the Horizontal System RIGOL Chapter 3 To Set the Horizontal System The contents of this chapter: Delayed Sweep Time Base Mode Horizontal Scale Horizontal Reference MSO2000A/DS2000A User’s Guide 3-1
RIGOL Chapter 3 To Set the Horizontal System Delayed Sweep Delayed sweep can be used to enlarge a length of waveform horizontally to view the waveform details. Press MENU in the horizontal control area (HORIZONTAL) at the front panel and press Delayed to enable or disable delayed sweep. In delayed sweep mode, the screen is divided into two display areas as shown in the figure below. Note: To enable delayed sweep, the current time base mode must be “Y-T” and the “Pass/Fail test” must be disabled.
Chapter 3 To Set the Horizontal System RIGOL The Waveform after Enlargement: The waveform in the lower part of the screen is the horizontally expanded waveform. Compared to the main time base, the delayed time base has increased the waveform resolution (as shown in the figure above). Note: The delayed time base should be less than or equal to the main time base and can be modified by rotating HORIZONTAL SCALE. Tip You can also press down HORIZONTAL to directly switch to delayed sweep mode.
RIGOL Chapter 3 To Set the Horizontal System Time Base Mode Press MENU in the horizontal control area (HORIZONTAL) at the front panel and then press Time Base to select the time base mode of the oscilloscope and the default is Y-T. Y-T Mode In this mode, the Y axis represents voltage and the X axis represents time. Note: Only when this mode is enabled can “Delayed Sweep” be turned on. In this mode, when the horizontal time base is greater than or equal to 200 ms, the instrument enters slow sweep mode.
Chapter 3 To Set the Horizontal System RIGOL According to sinθ=A/B or C/D (wherein, θ is the phase deviation angle between the two channels and the definitions of A, B, C and D are as shown in the figure above), the phase deviation angle is obtained, that is: θ=±arcsin (A/B) or ±arcsin( C/D) If the principal axis of the ellipse is within quadrant I and III, the phase deviation angle obtained should be within quadrant I and IV, namely within (0 to π/2) or (3π/2 to 2π).
RIGOL 4. Chapter 3 To Set the Horizontal System Observe the measurement result shown in the figure above. According to the measurement schematic diagram of the phase deviation (as shown in Figure 3-2), A/B (C/D) = 1. Thus, the phase deviation angle θ=±arcsin1=90°. Note: The maximum sample rate of X-Y mode is 1.0 GSa/s. Generally, longer sample waveform can ensure better display effect of Lissajous figure.
Chapter 3 To Set the Horizontal System RIGOL Roll Mode In this mode, the waveform scrolls from the right to the left to update the display. The range of horizontal scale adjustment is from 200.0 ms to 1.000 ks. Note: When Roll mode is enabled, the waveform “horizontal position”, “Delayed Sweep”, “Protocol Decoding”, “Pass/Fail Test”, “Measurement Range”, “Waveform Record”, “To Set the Persistence Time” and “To Trigger the Oscilloscope” are not available.
RIGOL Chapter 3 To Set the Horizontal System Horizontal Scale Horizontal scale, namely horizontal time base, refers to the time per grid in the horizontal direction on the screen and is usually expressed as s/div. Being similar to “Vertical Scale”, the horizontal scale can be adjusted in “Coarse” or “Fine” mode. Press MENU ScaleAdjust in the horizontal control area (HORIZONTAL) at the front panel to select the desired mode.
Chapter 3 To Set the Horizontal System RIGOL Horizontal Reference Horizontal reference is the reference position according to which the screen waveform is expanded or compressed horizontally when adjusting HORIZONTAL SCALE. In Y-T mode and the delayed sweep is disabled, press MENU HorRef in the horizontal control area (HORIZONTAL) at the front panel to select the desired reference mode and the default is “Center”.
Chapter 4 To Set the Sample System RIGOL Chapter 4 To Set the Sample System The contents of this chapter: Acquisition Mode Sample Mode Sample Rate LA Sample Rate LA Memory Depth Memory Depth Antialiasing MSO2000A/DS2000A User’s Guide 4-1
RIGOL Chapter 4 To Set the Sample System Acquisition Mode The acquisition mode is used to control how to generate waveform points from sample points. Press Acquire Acquisition and use to select the desired acquisition mode (the default is “Normal”), then press down the knob to select this mode. You can also press Acquisition continuously to switch the acquisition mode. Normal In this mode, the oscilloscope samples the signal at equal time interval to rebuild the waveform.
Chapter 4 To Set the Sample System RIGOL Figure 4-1 The Waveform before Average Figure 4-2 The Waveform after 256 Averages MSO2000A/DS2000A User’s Guide 4-3
RIGOL Chapter 4 To Set the Sample System Peak Detect In this mode, the oscilloscope acquires the maximum and minimum values of the signal within the sample interval to get the envelope of the signal or the narrow pulse of the signal that might be lost. In this mode, signal confusion can be prevented, but the noise displayed would be larger.
Chapter 4 To Set the Sample System RIGOL Sample Mode This oscilloscope only supports real-time sample. In this mode, the oscilloscope samples and displays waveform within a trigger event. The maximum real-time sample rate of the analog channels of MSO2000A/DS2000A is 2 GSa/s and the current sample rate is displayed under the Sa Rate menu. Tip Press RUN/STOP to stop the sample, the oscilloscope will hold the last display.
RIGOL Chapter 4 To Set the Sample System Sample Rate Sample refers to the process that the oscilloscope converts analog signals to digital signals at certain time interval and stores the signals in sequence. The sample rate is the reciprocal of the time interval. The maximum real-time sample rate of the analog channels of MSO2000A/DS2000A is 2 GSa/s.
RIGOL Chapter 4 To Set the Sample System from the sample data does not reflect all the actual signal information. Pulse disappeared LA Sample Rate LA sampling refers to the process that the oscilloscope samples the compared digital signal at a certain time interval. LA sample rate is the reciprocal of the time interval. For example, the oscilloscope with 500 MSa/s LA sample rate samples digital signals once every 2 ns. The maximum sample rate of the digital channels of the oscilloscope is 1 GSa/s.
RIGOL Chapter 4 To Set the Sample System Memory Depth Memory depth refers to the number of waveform points that the oscilloscope can store in a single trigger sample and it reflects the storage ability of the sample memory. MSO2000A/DS2000A provides 14 Mpts standard memory depth and up to 56 Mpts memory depth (option).
Chapter 4 To Set the Sample System RIGOL In “Auto” mode, the oscilloscope selects the memory depth automatically according to the current sample rate. At this point, you can adjust the memory depth indirectly by rotating HORIZONTAL SCALE to adjust the sample rate. LA Memory Depth Press Acquire, you can view the current memory depth of the digital channel in LA Mem Depth. The LA Memory Depth will change with the memory depth of the analog channel and can not be set separately.
Chapter 5 To Trigger the Oscilloscope RIGOL Chapter 5 To Trigger the Oscilloscope As for trigger, you set certain trigger condition according to the requirement and when a waveform in the waveform stream meets this condition, the oscilloscope captures this waveform as well as the neighbouring part and displays them on the screen. For digital oscilloscope, it displays waveform continuously no matter whether it is stably triggered, but only stable trigger can ensures stable display.
RIGOL Chapter 5 To Trigger the Oscilloscope Trigger Source Press MENU Source in the trigger control area (TRIGGER) at the front panel to select the desired trigger source. Signals input from CH1/CH2, the [EXT TRIG] connector or [LOGIC D0-D15] connector as well as the AC Line can all be used as trigger source. Analog channel input: Signals input from analog channels CH1 and CH2 can all be used as the trigger source. No matter whether the channel selected is enabled, the channel can work normally.
RIGOL Chapter 5 To Trigger the Oscilloscope Trigger Mode Trigger mode affects the way in which the oscilloscope searches for the trigger. The following is the schematic diagram of the acquisition memory. As shown in the figure below, the position of the trigger event is determined by the reference time point and the delay setting. Note: The acquisition memory of the oscilloscope is a cyclic buffer and the new data would overwrite the old data until the acquisition finishes.
RIGOL Chapter 5 To Trigger the Oscilloscope waveform is displayed. Otherwise, “AUTO” is displayed and unstable waveform is displayed. Note: When the horizontal time base setting is 50 ms/div or greater, you can also observe the input waveform without a trigger signal in this mode. Normal: After this mode is selected, the oscilloscope enters the wait-for-trigger state and starts searching for trigger signals that meet the specified condition.
Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Coupling Trigger coupling decides which kind of components will be transmitted to the trigger module. Please distinguish it with “Channel Coupling”. DC: allow DC and AC components into the trigger path. AC: block all the DC components and attenuate signals lower than 75 Hz. LF Reject: block the DC components and reject the low frequency components (lower than 75 kHz). HF Reject: reject the high frequency components (higher than 75 kHz).
RIGOL Chapter 5 To Trigger the Oscilloscope Trigger Holdoff Trigger holdoff can be used to stably trigger the complex waveforms. Holdoff time is the amount of time that the oscilloscope waits before re-arming the trigger module. The oscilloscope will not trigger even if the trigger condition is met during the holdoff time and will only re-arm the trigger module after the holdoff time expires.
Chapter 5 To Trigger the Oscilloscope RIGOL Noise Rejection Noise rejection can reject the high frequency noise in the signal and reduce the possibility of miss-trigger of the oscilloscope. Press MENU Setting Noise Reject in the trigger control area (TRIGGER) at the front panel to enable or disable noise rejection.
RIGOL Chapter 5 To Trigger the Oscilloscope Trigger Type MSO2000A/DS2000A provides various trigger types, including various serial bus triggers.
Chapter 5 To Trigger the Oscilloscope RIGOL Edge Trigger Trigger on the trigger threshold of the specified edge of the input signal. Trigger Type: Press Type, rotate to select “Edge” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen. For example, . The trigger type is edge trigger; the trigger source is CH1; the trigger level is 0.00 V. Source Selection: Press Source and select CH1, CH2, EXT, AC Line or any channel of D0-D15.
RIGOL Chapter 5 To Trigger the Oscilloscope LEVEL to modify the level. If the current trigger source is a digital channel, the trigger level value will be displayed at the upper right corner of the screen. If the current trigger source is an analog channel, an orange trigger level line and the trigger mark “ ” appear on the screen and move up or down with the rotation of the knob, while at the same time, the trigger level value (such as ) at the lower left corner of the screen also changes accordingly.
RIGOL Chapter 5 To Trigger the Oscilloscope Pulse Trigger In pulse trigger mode, the oscilloscope will trigger when the pulse width of the input signal satisfies the specified pulse width condition. Trigger Type: Press Type, rotate to select “Pulse” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen. For example, . The trigger type is pulse trigger; the trigger source is CH1; the trigger level is 0.00 V.
RIGOL Chapter 5 To Trigger the Oscilloscope the specified pulse width. : trigger when the positive pulse width of the input signal is lower than the specified pulse width. : trigger when the positive pulse width of the input signal is greater than the specified lower limit of pulse width and lower than the upper limit of pulse width. : trigger when the negative pulse width of the input signal is greater than the specified pulse width.
Chapter 5 To Trigger the Oscilloscope RIGOL Runt Trigger This trigger mode is used to trigger pulses that pass through one trigger level but fails to pass through the other trigger level as shown in the figure below. Figure 5-4 Runt Trigger Trigger Type: Press Type, rotate to select “Runt” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen. For example, .
RIGOL Chapter 5 To Trigger the Oscilloscope width. Press Lower Limit to set the minimum pulse width of runt trigger. The range available is from 2 ns to 4 s. <: trigger when the runt pulse width is lower than the upper limit of pulse width. Press Upper Limit to set the maximum pulse width of runt trigger. The range available is from 2 ns to 4 s. <>: trigger when the runt pulse width is greater than the lower limit and lower than the upper limit of pulse width.
Chapter 5 To Trigger the Oscilloscope RIGOL the “Low Level” changes accordingly and the “Up Level” remains unchanged. : adjust the upper and lower limits of the trigger level at the same time. During the adjustment, the “Up Level” and “Low Level” change accordingly. Trigger Mode: Press Sweep and select “Auto”, “Normal” or ”Single”. For the details, please refer to “Trigger Mode”. The corresponding status light of the current trigger mode turns on.
RIGOL Chapter 5 To Trigger the Oscilloscope Windows Trigger (Option) Windows trigger provides a high trigger level and a low trigger level. The oscilloscope triggers when the input signal passes through the high trigger level or the low trigger level. Trigger Type: Press Type, rotate to select “Windows” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen. For example, .
Chapter 5 To Trigger the Oscilloscope RIGOL the default is 1.00 us. Vertical Window and Trigger Level: Press Vertical to select the desired vertical window type. Rotate TRIGGER LEVEL to adjust the trigge level. For the detailed operation, please refer to “Vertical Window and Trigger Level”. Trigger Mode: Press Sweep and select “Auto”, “Normal” or ”Single”. For the details, please refer to “Trigger Mode”. The corresponding status light of the current trigger mode turns on.
RIGOL Chapter 5 To Trigger the Oscilloscope Nth Edge Trigger (Option) Trigger on the nth edge that appears after the specified idle time. For example, in the waveform shown in the figure below, the instrument should trigger on the second rising edge after the specified idle time (the time between two neighbouring rising edges) and the idle time should be set to P
Chapter 5 To Trigger the Oscilloscope RIGOL : trigger on the falling edge of the input signal when the voltage level meets the specified trigger level. Idle Time: Press Idle to set the idle time before the edge counting in Nth edge trigger. The range available is from 16 ns to 4 s. Edge Number: Press Edge to set the value of “N” in Nth edge trigger and the range available is from 1 to 65535. Trigger Mode: Press Sweep and select “Auto”, “Normal” or ”Single”.
RIGOL Chapter 5 To Trigger the Oscilloscope Slope Trigger Trigger on the positive or negative slope which satisfies the specified condition. Trigger Type: Press Type, rotate to select “Slope” and press down . At this point, the trigger setting information is displayed at the upper right corner of the screen. For example, . The trigger type is slope trigger; the trigger source is CH1; the difference between the upper limit of trigger level and the lower limit of trigger level is 1.24 V.
Chapter 5 To Trigger the Oscilloscope RIGOL : trigger when the negative slope time of the input signal is greater than the specified time. : trigger when the negative slope time of the input signal is lower than the specified time. : trigger when the negative slope time of the input signal is greater than the specified lower limit of time and lower than the specified upper limit of time.
RIGOL Chapter 5 To Trigger the Oscilloscope SlewRate = UpLevel - LowLevel Time When the Slope Condition is set to or , the current trigger level and slew rate range will be displayed at the lower left corner of the screen, as shown in figure 5-5 (b). The formula of slew rate range is: SlewRate = (a) UpLevel - LowLevel UpLevel - LowLevel ~ UpperLimit LowerLimit (b) Note: Under the “Slope” trigger menu, you can also press down the trigger level knob continuously to switch the vertical window.
Chapter 5 To Trigger the Oscilloscope RIGOL Video Trigger (HDTV Option) The waveform of video signal includes image signal and timing signal and different signals adopt different standards and formats. MSO2000A/DS2000A provides basic measurement function for video signals. The oscilloscope triggers on the standard video signal field or line of NTSC (National Television Standards Committee), PAL (Phase Alternating Line), SECAM (Sequential Couleur A Memoire) or HDTV (High Definition Television).
RIGOL Chapter 5 To Trigger the Oscilloscope Even Field: trigger on the rising edge of the first ramp waveform pulse in the even field. Video Standard: Press Standard to select the desired video standard. NTSC: the field frequency is 60 fields per second and the frame frequency is 30 frames per second. The TV sweep line is 525 with the even field goes first and the odd field follows behind. PAL: the frame frequency is 25 frames per second.
Chapter 5 To Trigger the Oscilloscope RIGOL Tips For a better observation of the waveform details in the vedio signal, you can set a larger memory depth first. In the trigger debugging process of vedio signals, the frequencies in different parts of the signal can be reflected by different brightnesses as RIGOL digital oscilloscopes provides the multi-level gray scale display function. Experienced users can quickly judge the signal quality and discover abnormalities.
RIGOL Chapter 5 To Trigger the Oscilloscope Pattern Trigger Identify a trigger condition by looking for a specified pattern. This pattern is a logical “AND” combination of the channels. Each channel can be set to high (H), low (L), don’t care (X), rising edge or falling edge. When a rising or falling edge is specified, the oscilloscope will trigger at the edge specified if the input signals for other channels are consistent with the pattern currently set.
Chapter 5 To Trigger the Oscilloscope RIGOL : set the pattern of the channel selected to “L”, namely the voltage level is lower than the threshold level of the channel. : set the pattern of the channel selected to “Don’t Care”, namely this channel is not used as a part of the pattern. When both the channels in the pattern are set to “Don’t Care”, the oscilloscope will not trigger. or : set the pattern to the rising or falling edge of the channel selected.
RIGOL Chapter 5 To Trigger the Oscilloscope Delay Trigger (Option) In delay trigger, you need to set signal source A and signal source B. The oscilloscope triggers when the time difference (△T) between the specified edges of source A (Edge A) and source B (Edge B) meets the preset time limit, as shown in the figure below. Note: Edge A and Edge B must be neighbouring edges. Edge A= Edge B= Source A Source B T Figure 5-8 Delay Trigger Trigger Type: Press Type to open the trigger type list.
Chapter 5 To Trigger the Oscilloscope RIGOL Source B: Press SourceB to select CH1, CH2 or any channel of D0-D15 as the trigger source of signal source B. Please refer to the introduction in “Trigger Source”. The signal source currently selected is displayed in the upper-right corner of the screen. Edge B: Press EdgeB to select the trigger edge type of signal source B in delay trigger. It can be set to the rising edge or falling edge.
RIGOL Chapter 5 To Trigger the Oscilloscope Trigger Setting: Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type. Note: Noise rejection is grayed out and disabled automatically when the signal source is set to any channel of D0-D15. Trigger Level: Press SourceA and use TRIGGER LEVEL to modify the trigger level of source A. Press SourceB and use TRIGGER LEVEL to modify the trigger level of source B.
RIGOL Chapter 5 To Trigger the Oscilloscope TimeOut Trigger (Option) In timeout trigger, the instrument triggers when the time interval (△T) from when the rising edge (or falling edge) of the input signal passes through the trigger level to when the neighbouring falling edge (or rising edge) passes through the trigger level is greater than the timeout time set, as shown in the figure below. Time Out< T T Time Out Figure 5-9 TimeOut Trigger Trigger Type: Press Type to open the trigger type list.
RIGOL Chapter 5 To Trigger the Oscilloscope Timeout Time: Press TimeOut to set the timeout time of timeout trigger and the range is from 16 ns to 4 s. Trigger Mode: Press Sweep to open the trigger mode list and select “Auto”, “Normal” or ”Single”. For the details, please refer to “Trigger Mode”. The corresponding status light of the current trigger mode turns on. Trigger Setting: Press Setting to set the trigger parameters (trigger holdoff and noise rejection) under this trigger type.
RIGOL Chapter 5 To Trigger the Oscilloscope Duration Trigger (Option) In duration trigger, the instrument identifies a trigger condition by looking for the duration of a specified pattern. This pattern is a logical “AND” combination of the two channels. Each channel can have a value of high (H), low (L) or don’t care (X). The instrument triggers when the duration (△T) of this pattern meets the preset time, as shown in the figure below.
RIGOL Chapter 5 To Trigger the Oscilloscope : set the pattern of the channel selected to “L”, namely the voltage level is lower than the threshold level of the channel. : set the pattern of the channel selected to “Don’t Care”, namely this channel is not used as a part of the pattern. When both the channels in the pattern are set to “Don’t Care”, the oscilloscope will not trigger. Trigger Condition: Press When to select the desired trigger condition.
RIGOL Chapter 5 To Trigger the Oscilloscope Setup/Hold Trigger In setup/hold trigger, the instrument triggers when the internal state of the setup or hold time relative to the clock edge is changed by the logic data input, namely trigger when the setup time (△T1) is less than the preset setup time or when the hold time (△T2) is less than the preset hold time, as shown in the figure below.
RIGOL Chapter 5 To Trigger the Oscilloscope Setup Type: Press SetupType to select the desired setup type. Setup: set the time that the data stays stable and constant before the clock edge appears. Press Setup to set the setup time and the range is from 2 ns to 1 s. Hold: set the time that the data stays stable and constant after the clock edge appears. Press Hold to set the hold time and the range is from 2 ns to 1 s.
Chapter 5 To Trigger the Oscilloscope RIGOL RS232 Trigger RS232 bus is a serial communication mode used in the data transimmision between PCs or between PC and terminal. In RS232 serial protocol, a character is transimitted as a frame of data which consists of 1 bit start bitm, 5~8 bits data bits, 1 bit check bit and 1~2 bits stop bit(s). Its format is as shown in the figure below.
RIGOL Chapter 5 To Trigger the Oscilloscope --press Stop Bit to select “1 bit” or “2 bit”; --press Even-Odd to select “None”, “Odd” or “Even”. The oscilloscope will determine error frame according to the preset parameters. Check Error: trigger when check error is detected. When this trigger condition is selected, press Even-Odd to select “Odd” or “Even”. The oscilloscope will determine check error according to the preset parameters. Data: trigger on the last bit of the preset data bits.
Chapter 5 To Trigger the Oscilloscope RIGOL I2C Trigger I2C is a 2-wire serial bus used to connect the microcontroller and its peripheral device and is a bus standard widely used in the microelectronic communication control field. The I2C serial bus consists of SCL and SDA. Its transmission rate is determined by SCL and its transmission data is determined by SDA, as shown in the figure below.
RIGOL Chapter 5 To Trigger the Oscilloscope Trigger Condition: Press When to select the desired trigger condition. Start: trigger when SDA data transitions from high to low while SCL is high. Restart: trigger when another start condition occurs before a stop condition. Stop: trigger when SDA data transitions from low to high while SCL is high. Missing ACK: trigger when the SDA data is high during any acknowledgement of SCL clock position.
Chapter 5 To Trigger the Oscilloscope RIGOL Trigger Mode: Press Sweep to open the trigger mode list and select “Auto”, “Normal” or ”Single”. For the details, please refer to “Trigger Mode”. The corresponding status light of the current trigger mode turns on. Trigger Setting: Press Setting to set the trigger parameters (noise rejection) under this trigger type. Note: Noise rejection is grayed out and disabled automatically when the signal source is set to any channel of D0-D15.
RIGOL Chapter 5 To Trigger the Oscilloscope SPI Trigger Trigger on the data pattern on the specified edge. When using SPI trigger, you need to specify the SCL and SDA data sources. Below is the sequence chart of SPI bus data transmission.
Chapter 5 To Trigger the Oscilloscope RIGOL Data Line Setting: Press Data to enter the data line setting munu.. Press Data Bits to set the number of bits of the serial data character string. It can be set to any integer between 4 and 32. Press CurrentBit to set the number of the data bit and the range is from 0 to (value specified in Data Bits – 1). Press Data to set the value of the current bit to H, L or X. Press AllBits to set all the data bits to the value specified in Data.
RIGOL Chapter 5 To Trigger the Oscilloscope Trigger Level: Press SCL and use TRIGGER LEVEL to modify the trigger level of the SCL channel. Press SDA and use TRIGGER LEVEL to modify the trigger level of the SDA channel. For the details, please refer to the introduction of “Trigger Level”. USB Trigger (Option) Trigger on the SOP, EOP, RC, Suspend and Exit Suspend of the data packet on the differential USB data cable (D+ and D-). This trigger supports USB Low Speed and Full Speed.
Chapter 5 To Trigger the Oscilloscope RIGOL SOP: trigger at the sync bit at the start of the data packet (SOP). EOP: trigger at the end of the SEO portion of the EOP of the data packet. RC: trigger when SEO is greater than 10 ms. Suspend: trigger when the idle time of the bus is greater than 3 ms. ExitSuspend: trigger when the bus exits from idle state for more than 10 ms. Trigger Mode: Press Sweep to open the trigger mode list and select “Auto”, “Normal” or ”Single”.
RIGOL Chapter 5 To Trigger the Oscilloscope CAN Trigger (Option) MSO2000A/DS2000A oscilloscope provides CAN bus trigger and decoding functions for better CAN bus analysis. Trigger on the start frame, end frame, specified frame type or error frame of the CAN signal. When using CAN trigger, you need to specify the signal source, signal rate and trigger signal type of the CAN signal. The figure below shows the standard and expanded formats of CAN bus data frame.
Chapter 5 To Trigger the Oscilloscope RIGOL current trigger source is displayed at the upper right corner of the screen. Signal Type: Press Signal Type to select the desired signal type. Rx: receiving signal on the CAN signal line. Tx: transmission signal on the CAN signal line. CAN_H: actual CAN_H bus signal. CAN_L: actual CAN_L bus signal. Differential: CAN differential bus signal connected to the analog channel via the differential probe.
RIGOL Chapter 5 To Trigger the Oscilloscope Baud: Press Baud to set the CAN baud to match the CAN bus signal and the values available are 10 kb/s (default), 20 kb/s, 33.3 kb/s, 50 kb/s, 62.5 kb/s, 83.3 kb/s, 100 kb/s, 125 kb/s, 250 kb/s, 500 kb/s, 800 kb/s, 1 Mb/s or user. After “User” is selected, press User and use or the nevigation knob to input the desired rate within 1 kb/s and 1 Mb/s. Sample Point: Sample point is a point within a bit’s time. The oscilloscope samples the bit level at this point.
RIGOL Chapter 5 To Trigger the Oscilloscope Trigger Output Connector The trigger output connector at the rear panel can output trigger signals determined by the current setting. Trigger Output Connector Press Utility AuxOutput to select “TrigOut”. When the oscilloscope is triggered, it will output a trigger signal determined by the current trigger setting through the [Trigger Out] connector.
Chapter 6 MATH and Measurements RIGOL Chapter 6 MATH and Measurements MSO2000A/DS2000A can make math operation, auto measurement and cursor measurement on sampled and displayed data.
RIGOL Chapter 6 MATH and Measurements Math Operation MSO2000A/DS2000A can realize various math operations (including: addition (A+B), subtraction (A-B), multiplication (AxB), division (A÷B), FFT, logic operation and advanced operation) of waveforms between channels. The results of math operation also allow further measurement. Press MATH Operate in the vertical control area (VERTICAL) at the front panel to select the desired operation function.
Chapter 6 MATH and Measurements RIGOL HORIZONTAL POSITION and HORIZONTAL SCALE can also be used to adjust the horizontal position and scale of the operation results. Multiplication Multiply the waveform voltage values of signal source A and B point by point and display the results. Press MATH Operate to select “A x B”: Press Source A and Source B to select the desired channels (CH1 or CH2). Press and use to adjust the vertical position of the operation results.
RIGOL Chapter 6 MATH and Measurements FFT FFT is used to quickly perform Fourier transform on specified signals and transform time domain signals to frequency domain signals. MSO2000A/DS2000A oscilloscope provides FFT operation function which enables users to observe the time domain waveform and spectrum of the signal at the same time.
Chapter 6 MATH and Measurements RIGOL MSO2000A/DS2000A provides four kinds of FFT window functions which have different characteristics and are applicable to measure different waveforms. You need to select the window function according to the waveform to be measured and its characteristics. Press Window to select the desired window function and the default is “Rectangle”. Table 6-1 Window Functions 3.
RIGOL Chapter 6 MATH and Measurements Note: In FFT mode and when MATH is the active channel, you can also press HORIZONTAL SCALE to switch between “Split” and “Full Screen”. 4. Set Horizontal Position and Horizontal Scale The horizontal axis of the FFT operation results represents frequency and its unit is Hz. Use HORIZONTAL POSITION and HORIZONTAL SCALE to set the horizontal position and horizontal scale of the FFT frequency domain waveform respectively.
Chapter 6 MATH and Measurements RIGOL Logic Operation Perform logic operation on the waveform voltage values of the specified sources point by point and display the results. In operation, when the voltage value of the source channel is greater than the threshold of the corresponding channel, it is regarded as logic “1”; otherwise logic “0”. The following common logic operation expressions are provided.
RIGOL Chapter 6 MATH and Measurements Press MATH Operate to select “Logic”: Press Formula to select the desired logic formula and the default is “AND”. Press Source A and Source B to select the desired channels (CH1, CH2 or any channel of D0-D15). 6-8 Press and use to adjust the vertical position of the operation results. Press and use to adjust the vertical scale of the operation results. Press Invert to turn the inverted display of the operation results on or off.
Chapter 6 MATH and Measurements RIGOL Advanced Operation MSO2000A/DS2000A provides advanced operation function that allows users to define operation functions. Press MATH Operate “Advanced” Expression “ON” and the editing window as shown in the figure below is displayed.
RIGOL 3. Chapter 6 MATH and Measurements Function The functions of the functions are listed in the following table. Note: The left brackets “(” here are only used to facilitate your entry and they are not a part of the function names. Table 6-3 Functions 4. 6-10 Name Function Intg( Calculate the integral of the selected source. You can use integral to measure the area under a waveform or the pulse energy. Diff( Calculate the discrete time differentiate of the selected source.
Chapter 6 MATH and Measurements Mantissa: Exponent: 5. RIGOL 6.1074 8 Operator The functions of the operators are listed in the following table.
RIGOL Chapter 6 MATH and Measurements Auto Measurement MSO2000A/DS2000A provides auto measurements of 29 waveform parameters and the statistics and analysis of the measurement results. What’s more, you can also use the frequency counter to realize more precise frequency measurement.
Chapter 6 MATH and Measurements RIGOL One-key Measurement of 29 Parameters Press MENU at the left of the screen to turn on the measurement menu of the 29 parameters and then press the corresponding menu softkey to quickly realize “One-key” measurement. The measurement result will be displayed at the bottom of the screen. Note: Press Measure Source to select the current measurement source.
RIGOL Chapter 6 MATH and Measurements Time Parameters FallTime RiseTime Threshold Upper Limit +Width Threshold Middle Value -Width Threshold Lower Limit Period Figure 6-3 Time Parameters 1. 2. 3. 4. 5. 6. 7. 8. Period: defined as the time between the middle threshold points of two consecutive, like-polarity edges. Frequency: defined as the reciprocal of period. Rise Time: the time for the signal amplitude to rise from the threshold lower limit to the threshold upper limit.
RIGOL Chapter 6 MATH and Measurements Delay and Phase Period Source 1 Delay Source 2 Figure 6-4 Delay and Phase Source 1 and Source 2 can be CH1, CH2 or any channel of D0-D15. You can set them via Measure Setting Type ”Delay” (“Phase”). For the setting method, please refer to the detailed introduction in “Measurement Setting”. 1. Delay 1 →2 : the time difference between the rising edges of source 1 and 2. source 2.
RIGOL 8. Chapter 6 MATH and Measurements below. Phase 1 →2 : phase difference calculated according to “Delay 1 →2 ” and the period of source 1, expressed in degree. The calculation formula is as shown below.
Chapter 6 MATH and Measurements RIGOL Voltage Parameters Overshoot Vmax Vtop Vamp Vpp Vbase Vmin Preshoot 1. 2. 3. 4. 5. 6. 7. Figure 6-5 Voltage Parameters Vmax: the voltage value from the highest point of the waveform to the GND. Vmin: the voltage value from the lowest point of the waveform to the GND. Vpp: the voltage value from the highest point to the lowest point of the waveform. Vtop: the voltage value from the flat top of the waveform to the GND.
RIGOL Chapter 6 MATH and Measurements 10. Overshoot: the ratio of the difference of the maximum value and top value of the waveform to the amplitude value. 11. Preshoot: the ratio of the difference of the minimum value and base value of the waveform to the amplitude value. Other Parameters 1. Area: the area of the whole waveform within the screen and the unit is voltage-second.
Chapter 6 MATH and Measurements RIGOL Frequency Counter Measurement The hardware frequency counter supplied with MSO2000A/DS2000A oscilloscope can make more precise measurement of the input signal frequency. Press Measure Counter to select CH1, CH2 or any channel of D0-D15 as the measurement source. Besides, the signal input from the [EXT TRIG] connector can also be used as measurement source.
RIGOL Chapter 6 MATH and Measurements Measurement Setting 1. Source Selection Press Measure Source to select the desired channel for measurement (CH1, CH2, MATH or any channel of D0-D15). The color of the parameter icons under MENU at the left of the screen will change with the source selected. 2. Measurement Range Press Measure Range Range to select “Screen” or “Cursor” for measurement. When “Cursor” is selected, two cursor lines appear on the screen.
Chapter 6 MATH and Measurements RIGOL “Delay 1 →2 ”, “Delay 1 →2 ” and “Delay 1 →2 ”. Press Measure Setting Type “Delay” and then press SourceA and SourceB to set the two channel sources (CH1, CH2 or any channel of D0-D15) of delayed measurement respectively. Note: When SourceA is set to any analog channel (or digital channel), SourceB can only be set to a analog channel (or digital channel). 4.
RIGOL Chapter 6 MATH and Measurements To Clear the Measurement If you have currently enabled one or more items in the 29 measurement parameters, you can “Delete” or “Recover” the last five parameters or “Delete” or “Recover” all the measurement items enabled. Note: The last five parameters are determined according to the order in which you turned them on and they will not change as you delete one or more measurement items.
RIGOL Chapter 6 MATH and Measurements All Measurement All measurement could measure all the time and voltage parameters (each measurement source has 21 items, measurements can be performed on CH1, CH2 and MATH at the same time) of the current measurement source and display the results on the screen. Press Measure Display All to enable or disable the all measurement function. Press Measure Source and use to select the channel(s) to be measured (CH1, CH2 and MATH).
RIGOL Chapter 6 MATH and Measurements Statistic Function Make statistic and display the current, average, minimum (or standard deviation) and maximum (or count) values of at most 5 measurement items that are turned on last. Press Measure Statistic to turn the statistic function on or off. When the statistic function is enabled, press Mode to select “Extremum” or “Difference” measurement. When “Extremum” is selected, minimum and maximum values are displayed.
Chapter 6 MATH and Measurements RIGOL Cursor Measurement Before making cursor measurements, connect the signal to the oscilloscope and acquire stable display. Cursor measurement can measure the X axis values (usually Time) and Y axis values (usually Voltage) of the selected waveform. All the waveform parameters in “Auto Measurement” can be measured through cursor. Press Cursor Mode at the front panel and use to select the desired cursor mode (the default is “OFF”) and then press down the knob.
RIGOL Chapter 6 MATH and Measurements A->Y: the Y value at cursor A. Y value takes the channel GND of CH1 as reference. Its unit is the same as that of the current signal source. B->X: the X value at cursor B. X value takes the trigger position as reference. Its unit is “s” or “Hz” (when measuring FFT waveform). B->Y: the Y value at cursor B. Y value takes the channel GND of CH1 as reference. Its unit is the same as that of the current signal source.
Chapter 6 MATH and Measurements RIGOL 1. Select Display Mode Press DisplayMode to select “X”, “Y” or “X-Y”. The X cursors are a pair of vertical solid/dotted lines and are usually used to measure time parameters. The Y cursors are a pair of horizontal solid/dotted lines and are usually used to measure voltage parameters. The X-Y cursors are a pair of vertical solid/dotted lines and a pair of horizontal solid/dotted lines and are usually used to measure time and voltage parameters at the same time.
RIGOL Chapter 6 MATH and Measurements △X are in “s” and 1/△X is in “Hz”. Hz: when this unit is selected, in the measurement results, A->X, B->X and △X are in “Hz” and 1/△X is in “s”. °: when this unit is selected, in the measurement results, A->X, B->X and △X are in “°”. At this point, A->X, B->X and △X will change to “0°”, “360°” and “360°” respectively when you press SetCursor, no matter where cursor A and B are currently located.
Chapter 6 MATH and Measurements 6. RIGOL Adjust the vertical positions of the cursors: When the display mode is “Y” or “X-Y” (press SelectCursor to select “Y”) mode, press CursorA and CursorB and use to adjust the vertical positions of cursor A and cursor B respectively. Press CursorAB and use to adjust the vertical positions of cursor A and B at the same time. During the adjustment, the measurement result will change accordingly. The adjustable range is limited within the screen.
RIGOL Chapter 6 MATH and Measurements Track Mode In this mode, one or two cursors will appear. You can adjust the two cursors (cursor A and cursor B) to measure the X and Y values on two different sources respectively. The points being measured on cursor A and B are marked by an orange rectangle and rhombus respectively. When the cursors are moved horizontally, the marks will position on the waveform automatically.
Chapter 6 MATH and Measurements RIGOL namely do not use cursor A or cursor B. 2. Adjust Cursor Position Adjust cursor A: press CursorA and use to adjust the position of cursor A. During the adjustment, the measurement result will change accordingly. The adjustable range is limited within the screen. Adjust cursor B: press CursorB and use to adjust the position of cursor B. During the adjustment, the measurement result will change accordingly. The adjustable range is limited within the screen.
RIGOL Chapter 6 MATH and Measurements Figure 6-9 Cursor Track (Before Horizontal Expansion) Figure 6-10 Cursor Track (After Horizontal Expansion) 6-32 MSO2000A/DS2000A User’s Guide
Chapter 6 MATH and Measurements RIGOL Auto Mode In this mode, one or more cursors will appear. You can use auto cursor measurement to measure any of the 29 waveform parameters. Before using this mode, you need to at least enable one auto measurement parameter and the number of cursors will change with the measurement parameter enabled.
RIGOL Chapter 6 MATH and Measurements X-Y Mode X-Y mode is only available when the main timebase is “X-Y” mode. In this mode, two pairs of cursors will appear. You can adjust the cursor to measure X and Y values of the crossing points of the two pairs of cursors. Press MENU at the horizontal control are (HORIZONTAL) at the front panel and then press Time Base to select “X-Y”. Press Cursor Mode “X-Y” to enable the X-Y mode cursor measurement function.
Chapter 7 Digital Channel RIGOL Chapter 7 Digital Channel MSO2000A series mixed signal digital oscilloscope provides 2 analog channels as well as 16 digital channels. For the digital channels, the oscilloscope compares the voltages acquired in each sample with the preset logic threshold. If the voltage of the sample point is greater than the logic threshold, it will be stored as logic 1; otherwise, it will be stored as logic 0.
RIGOL Chapter 7 Digital Channel To Select the Digital Channel Press LA CH/Group, rotate the multifunction knob to select any of the digital channels or user-defined digital channel groups. You can also press CH/Group continuously or to switch among the options in order. D0-D15: select any channel of D0-D15. The channel label and waveform corresponding to the channel selected are displayed in red. Group1-Group4: select any of the user-defined groups Group1-Group4.
Chapter 7 Digital Channel RIGOL To Turn on/off the Digital Channel Press LA On/Off to enter the digital channel/group “On/Off” setting menu. Press D7-D0 to turn on or off channels D7-D0 at the same time. Press D15-D8 to turn on or off channels D15-D8 at the same time. Press Sel CH to turn on the channel list. Rotate to select any channel and press to turn on or off the channel. You can also press Sel CH continuously to turn on or off the selected channel.
RIGOL Chapter 7 Digital Channel Group Set Press LA Group Set to enter the user-defined group setting menu. You can perform group and ungroup operations on the 16 digital channels. Group: The group operations of Group1-Group4 are the same and Group1 is taken asaneaxample for illustration. Press Group1 to turn on the channel list (the status label is located at the left side of each channel), rotate to select the channel to be added into Group1 and press or Group1 to add the channel selected to Group1.
Chapter 7 Digital Channel RIGOL To Set the Waveform Display Size Press LA WaveSize, rotate to select the desired display size and the press down . You can also press WaveSize continuously to switch among S (small), M (medium) and L (large). Note: L (large) can only be used when the number of channels currently turned on is no more than 8. Reorder Setting Press LA ReOrder to select the waveform order mode on the screen of the channels currently selected.
RIGOL Chapter 7 Digital Channel To Use the Digital Bus Users can display the three groups of channels (D7-D0, D15-D8 and D15-D0) as digital buses according to their needs. Each bus value will be displayed at the bottom of the screen in data or graph form. You can create up to 2 buses. Press LA Bus to enter the digital bus setting menu. Press Bus to select “BUS1” or “BUS2”. Press BusStatus to turn on or off the digital bus. Note: The bus status of BUS1 and BUS2 can be set to “ON” at the same time.
RIGOL Chapter 7 Digital Channel Press Format to set the display format of the digital bus to Hex, Decimal, Binary or ASCII. Press Jitter Reject to turn on the jitter rejection function. Jitter refers to the short-term deviation of the signal relative to its ideal time position at certain time point, as shown as T1 and T2 in Figure 7-1.
RIGOL Chapter 7 Digital Channel set a label. Press Input to enter the label input interface to input a label manually. For the input method, refer to the introduction in “Channel Label”. Digital Channel Delay Calibration When using the oscilloscope for actual measurement, the transmission delay of the probe cable may cause relatively larger error (zero offset). Zero offset is defined as the offset of the crossing point of the waveform and threshold level line relative to the trigger position.
Chapter 8 Protocol Decoding RIGOL Chapter 8 Protocol Decoding Protocol analysis can be used by users to discover errors, debug hardware and accelerate development easily, so as to guarantee quick and high-quality accomplishment of projects. Protocol decoding is the basis of protocol analysis. Only protocol analyses with correct protocol decoding are acceptable and only correct protocol decoding can provide more error information.
RIGOL Chapter 8 Protocol Decoding Parallel Decoding Parallel bus consists of clock line and data line. As shown in the figure below, CLK is the clock line, while Bit0 and Bit1 are the 0 bit and 1st bit on the data line respectively. The oscilloscope will sample the channel data on the rising edge, falling edge or the rising/falling edge of the clock and judge each data point (logic “1” or logic “0”) according to the preset threshold level.
Chapter 8 Protocol Decoding RIGOL default is 0 and the range available is from 0 to (the bus bits-1). For example, when the bus bits is 18, the range available is 0, 1…17. Next, press Channel to specify a channel source for the bit currently selected in CurrentBit. The channel source can be CH1, CH2 or any channel of D0-D15. 3. Analog Channel Threshold Setting Press Threshold to turn the threshold setting menu on. Channel Select the channel (CH1 or CH2) that needs to set a threshold.
RIGOL Chapter 8 Protocol Decoding Figure 8-2 Parallel Event Table At this point, “Table” is selected in the View menu automatically and the menu is grayed out. The decoded data is displayed in table format following the time order. If an USB storage device is currently connected to the instrument, press Export to export the data table to the external USB storage device in CSV format.
RIGOL Chapter 8 Protocol Decoding RS232 Decoding (Option) RS232 serial bus consists of the transmitting data line (TX) and the receiving data line (RX). Rx Tx DeviceA DeviceB Tx Rx Figure 8-3 RS232 Serial Bus Schematic Diagram The industry standard of RS232 uses “Negative Logic”, namely high level is logic “0” and low level is logic “1”.
RIGOL Chapter 8 Protocol Decoding Start Bit: represent when the data starts outputting. Data Bits: represent the number of data bits actually contained in each frame of data. Check Bit: used to check the correctness of the data transmission. Odd Checkout: the total number of “1” in the data bit and check bit is an odd. For example, when 0x55 (01010101) is sent, a 1 needs to be filled in the check bit to make the number of 1 be an odd.
Chapter 8 Protocol Decoding RIGOL 5. Baud Rate Setting Press Baud to select the desired baud rate and the default is 9600 bps. If “User” is selected, press Setup and rotate or the navigation knob to set a desired baud rate. The range is from 50 to 20000000. 6. Data Packet Setting As mentioned before, in RS232, you need to set the start bit, data bits, check bit (optional) and stop bit of each frame of data. “Start Bit” is specified by the “Polarity Setting”.
RIGOL Chapter 8 Protocol Decoding Figure 8-5 RS232 Event Table At this point, “Table” is selected in the View menu automatically and the menu is grayed out. The event data is displayed in table format following the time order. If error occurs during the decoding, the corresponding error information is displayed. If a USB storage device is currently connected to the instrument, press Export to export the data table to the external USB storage device in CSV format. 9.
Chapter 8 Protocol Decoding RIGOL displayed. For example, when the transmitting terminal is set to none check and the decoder is set to odd check, the following check error occurs. (00000100,LSB) The check bit detected is 1 Wherein, there are odd number (1) of 1 in the 8-bit data 00000100 and the check bit should be 0; but the check bit detected on the TX is 1, thus check error occurs. After the decoder is set to none check, the decoding becomes normal.
RIGOL Chapter 8 Protocol Decoding I2C Decoding (Option) I2C serial bus consists of the clock line (SCLK) and the data line (SDA). SCLK: sample the SDA on the clock rising edge or falling edge. SDA: denote the data channel. Vcc A2 SCLK SDA Device A1 A0 Host SCLK SDA Figure 8-6 I2C Serial Bus Press Decode1 Decode to select “I2C” and open the I2C decoding function menu. 1. SCLK Setting Press SCLK to select any channel (CH1, CH2 or any channel of D0-D15) as the clock channel.
Chapter 8 Protocol Decoding RIGOL vertical position and scale, and is from (-5*vertical scale - vertical position) to (5*vertical scale - vertical position). If the data channel is set to any channel of D0-D15, SDAThreshold will be hidden automatically. 3. Copy Trigger Press CopyTrig to config the decoder parameter from the corresponding trigger. In I2C decoding, the function can only be enabled when the trigger type is I2C trigger, and you can copy the parameters of clock channel and data channel.. 4.
RIGOL Chapter 8 Protocol Decoding Figure 8-7 I2C Event Table (Table) Figure 8-8 I2C Event Table (Details) 6. 8-12 Address Information during Decoding In I2C bus, the front part of each frame of data contains the address information and blue patches are used to represent address ID. In the ID, “Write” is used to represent writing address and “Read” is used to represent reading address. Press Include R/W. When “Open” is selected, “AddrBits” will include the “R/W” bit as a part of the address value.
RIGOL Chapter 8 Protocol Decoding Address Information Figure 8-9 Address Information during Decoding 7. Error Expressions during Decoding When the ACK (ACKnowledge Character) is not met, the red error marks as shown in the figure below will be displayed. Note: The red mark is displayed in different modes according to the horizontal time base setting; when the horizontal time base is relatively smaller, displayed.
RIGOL Chapter 8 Protocol Decoding SPI Decoding (Option) SPI bus is based on the master-slave configuration and usually consists of chip select line (CS), clock line (SCLK) and data line (SDA). Wherein, the data line includes MISO and MOSI. SCLK: sample the SDA on the clock rising edge or falling edge. SDA: denote the data channel. SCLK Master MOSI MISO CS Slave SCLK SDA (MISO/MOSI) Figure 8-11 SPI Serial Bus Press Decode1 Decode to select “SPI” and open the SPI decoding function menu. 1.
Chapter 8 Protocol Decoding RIGOL Press Polarity to set the polarity of the SDA data line to (high) or (low). If the data channel is set to CH1 or CH2, press Threshold to set the threshold of the data channel. The range of the threshold is decided by the vertical position and scale, and is from (-5 * vertical scale - vertical position) to (5 * vertical scale - vertical position). If the data channel is set to any channel of D0-D15, Threshold will be hidden automatically. 3.
RIGOL Chapter 8 Protocol Decoding the event table interface as shown in the figure below. Figure 8-12 SPI Event Table At this point, “Table” is selected in the View menu automatically and the menu is grayed out. The decoded data is displayed in table format following the time order. If a USB storage device is currently connected to the instrument, press Export to export the data table to the external USB storage device in CSV format. 9.
Chapter 8 Protocol Decoding RIGOL CAN Decoding (Option) Press Decode1 Decode and select “CAN” to open the CAN decoding function menu. 1. Source Press Source and select any channel (CH1, CH2 or any channel of D0-D15) as the source channel. If the source channel is set to CH1 or CH2, press Threshold to set the threshold of the source channel.
RIGOL Chapter 8 Protocol Decoding 1bit 60% 70% 80% Figure 8-14 Sample Position Schematic Diagram 5. Copy Trigger Press CopyTrig to config the decoder parameter from the corresponding trigger. In CAN decoding, the function can only be enabled when the trigger type is CAN trigger, and you can copy the parameters of source, signal type, baud and sample point. 6. Display-related Setting Press Format to set the bus display format to Hex, Decimal, Binary or ASCII.
Chapter 8 Protocol Decoding RIGOL Figure 8-15 CAN Event Table (Table) Figure 8-16 CAN Event Table (Details) MSO2000A/DS2000A User’s Guide 8-19
RIGOL 8. Chapter 8 Protocol Decoding Decoded CAN Data Interpretation Frame ID: displayed as hex digits in blue. Data Length code (DLC): displayed as a chartreuse patch. Data Frame: displayed as green patches if data is successfully decoded. The frames appear as red patches if the data frame is lost. Cyclic Redundancy Check (CRC): displayed in a light blue patch when valid and red error mark is displayed when error occurs.
Chapter 9 Reference Waveform RIGOL Chapter 9 Reference Waveform The contents of this chapter: To Enable REF Function To Set the Color To Select REF Source To Save REF Waveform To Adjust REF Waveform Display To Export to Internal or External Memory To Import from Internal or External Memory MSO2000A/DS2000A User’s Guide 9-1
RIGOL Chapter 9 Reference Waveform To Enable REF Function Press REF in the vertical control area (VERTICAL) at the front panel to enable the REF function. Note: When the time base is in X-Y mode, REF function cannot be enabled. MSO2000A/DS2000A provides 10 reference waveform channels. Press Channel, rotate to select the desired reference channel and press down or Channel to turn the selected channel on or off.
Chapter 9 Reference Waveform RIGOL To Save REF Waveform After the REF source is selected, press Save to save the waveform (screen region) in the specified source to internal memory as reference waveform and display it on the screen. Note: This operation only saves the reference waveform in the volatile memory and the waveform will be cleared at power-off. To Adjust REF Waveform Display You can adjust the vertical scale and vertical position of the reference waveform specified in Current.
RIGOL Chapter 9 Reference Waveform To Import from Internal or External Memory Users can also import the reference waveform stored in the internal memory or external USB storage device to the instrument. Press Import to enter the file recall interface. Please refer to the relative descriptions in “Store and Recall” to import the reference waveform to the instrument.
Chapter 10 Pass/Fail Test RIGOL Chapter 10 Pass/Fail Test During the product design and manufacture process, you usually need to monitor the change of the signal or judge whether the product is up to standard. The pass/fail test function of MSO2000A/DS2000A oscilloscope can fulfill this task perfectly.
RIGOL Chapter 10 Pass/Fail Test To Enable Pass/Fail Test Press Utility Pass/Fail Enable to select “ON”. The blue shadow area denotes the failure area and if the waveform measured at certain time point during the measurement process passes through the failure area, it will be treated as failed frame. Press Operate to select “ ” to start testing and select “■” to stop testing. Note: The pass/fail test function is not available in the following status.
RIGOL Chapter 10 Pass/Fail Test horizontal and vertical adjustment ranges are 0.02 div to 4.0 div and 0.04 div to 5.12 div respectively. Figure 10-1 Creat Mask To Set Test Results Output Type You can set the output type of the test results referring to the following methods. 1. The Statistic Message Press Message to select “ON” or “OFF”.
RIGOL Chapter 10 Pass/Fail Test waveforms are detected. The test results on the screen will update continuously and a pulse will be output from the [Trigger Out] connector at the rear panel each time a failed waveform is detected. 3. Beeper on Fail Press Output to select “Fail” or “Fail+ ”. Fail: when a failed waveform is detected, the beeper does not sound. 4. Fail+ : when a failed waveform is detected, the beeper sounds (not related to the on/off state of the sound).
Chapter 11 Waveform Record RIGOL Chapter 11 Waveform Record Waveform record can record the waveforms of the analog channels (CH1 and CH2) and the digital channels (D0-D15). In record constant on (open) mode, the oscilloscope can record the input waveform continuously until users press RUN/STOP. Waveform playback and analysis can provide better waveform analysis effect. Note: 1. The horizontal time base must be set to Y-T mode during waveform record. 2.
RIGOL Chapter 11 Waveform Record Waveform Record When the waveform record function is turned on, waveforms from all the channels currently turned on will be recorded. Press Utility Record Mode and use to select “Record” to open the waveform record operation menu. 1. End Frame Press End Frame and use or the navigation knob to set the desired number of frames to be recorded. The available range is from 1 to the maximum number of frames that can be recorded currently (displayed in Total Frames). 2.
RIGOL Chapter 11 Waveform Record Note[1]: Memory depth=Sample rate×Horizontal time base×Number of grids horizontally on the screen; wherein, for MSO2000A/DS2000A, “Number of grids horizontally on the screen” is 14. Therefore, the maximum end frame of waveform record is also related to the “Sample rate” and “Horizontal time base”. Table 11-1 Memory Depth and Total Frames Memory Depth Maximum End Frame Auto 65000 14 k points 8128 140 k points 508 1.
RIGOL Chapter 11 Waveform Record Record Constant On MSO2000A/DS2000A provides record “Open” (record constant on) mode for waveform record. When this mode is enabled, the oscilloscope records the waveform continuously according to the current memory depth. The length of the waveform recorded is limited by the memory depth (sample rate and horizontal time base) and the waveform data recorded before will be overwritten by the waveform currently recorded.
Chapter 11 Waveform Record 3. 4. RIGOL Press RUN/STOP to stop the record. Analyze the waveform recorded using waveform analysis and as shown in the figure below, the runt pulses are captured.
RIGOL Chapter 11 Waveform Record Waveform Playback Waveform playback can play back the waveforms currently recorded. After recording the waveforms, press Utility Record Mode and use to select “Play back”. At this point, the information as shown in the figure below is displayed at the upper right corner of the screen, indicating the specific frame currently displayed on the screen. During the playback, this value would change continuously.
RIGOL Chapter 11 Waveform Record 5. End Frame Press this softkey and use to set the end frame of the playback. The default is the total number of frames of the waveform recorded. 6. Playback Operation Waveform playback can be realized through the menu or the shortcut buttons at the front panel. Menu Front Panel Press Operate and select ; the backlight flashes, indicating Press that the playback is started. “ ” to start playing back. Press Operate and select “ ” to pause the playback.
RIGOL Chapter 11 Waveform Record Waveform Analysis This function is used to analyze the recorded waveform. Press Utility Record Mode and use to select “Analyze” to open the waveform analysis menu. At this point, the screen is divided into two display areas as shown in the figure below. Settings and Results Templet Display Current Frame Figure 11-3 Waveform Analysis Please refer to the following explanations to set the waveform analysis parameters. 1.
Chapter 11 Waveform Record 3. RIGOL Start Before performing the analysis, please set the templet by referring to “Analysis Based on Trace” or create the mask by referring to “Analysis Based on Pass/Fail Mask” according to the analysis mode currently selected. Then, press Start to enable waveform analysis. Note: During the analysis process, the progress bar is displayed and the parameters cannot be modified.
RIGOL Chapter 11 Waveform Record whose difference value is equal to or larger than the threshold selected as an error frame and the corresponding “CurFrame Diff” is 100%; otherwise, the frame is judged as correct and the “CurFrame Diff” is 0%. Note: There are only two “CurFrame Diff” values (100% and 0%) under the analysis based on pass/fail mask. 4. Cancel During the analysis, users can press Cancel to stop the analysis and press Start again to restart the analysis. 5.
Chapter 11 Waveform Record 9. RIGOL Threshold: press this key to set the threshold of waveform analysis and the range is from 1% to 99%. The threshold is used to judge whether the frame is an error frame. A frame is recognize as an error one if the (relative) difference value between the frame and the templet is equal to or larger than the threshold currently set. Time Tag The time tag is used to display the absolute recording time of each frame of the waveform currently recorded.
RIGOL Chapter 11 Waveform Record Analysis Based on Trace Press Analyze and select “Trace”. Then, set the templet used in analysis based on trace through the method below. 1. Trace Press Trace to select the creation method of analysis templet. Current Frame: select the current frame as the analysis templet. Average: select the average of the current data frame as the analysis templet. 2. Setup Template Press SetupTemplate to configure the templet immediately.
Chapter 11 Waveform Record RIGOL Analysis Based on Pass/Fail Mask Press Analyze and select “Pass/Fail”. Then, set the templet used in analysis based on Pass/Fail mask through the method below. Press Range to open the following setting menus. 1. Range Press this softkey to select “Screen” or “Cursor” for the analysis (the default is “Screen”). When “Cursor” is selected, two gray cursor lines appear on the screen.
RIGOL Chapter 11 Waveform Record 4. Create Mask Mask is the templet used in waveform analysis. Press Create to immediately apply the mask (X Mask and Y Mask) currently created. 5. Save or Recall the Mask Users can store the current test mask in “*.pf” format into the internal memory or external USB storage device or load the test mask file stored in the internal memory or external USB storage device into the instrument. Press Save to enter the file store interface.
Chapter 12 Display Control RIGOL Chapter 12 Display Control You can set the type, persistence time and brightness of waveform display as well as the grid type, grid brightness of the screen display and the menu display time.
RIGOL Chapter 12 Display Control To Select the Display Type Press Display Type to set the waveform display type to “Vectors” or “Dots”. Vectors: the sample points are connected by lines and displayed, as shown in the left figure below. Normally, this mode can provide the most vivid waveform to view the steep edge of the waveform (such as square waveform). Dots: display the sample points directly, as shown in the right figure below.
Chapter 12 Display Control RIGOL 2. Specific Values Enable to observe glitch that changes relatively slowly or glitch with low occurrence probability. The persistence time can be set to 50 ms, 100 ms, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s or 20 s. 3. Infinite In this mode, the oscilloscope displays the waveform newly acquired without clearing the waveforms acquired formerly.
RIGOL Chapter 12 Display Control To Set the Waveform Intensity Press Display WaveIntensity or turn when the menu is hidden to adjust the waveform brightness of the analog channel. The default is 50% and the range available is from 0% to 100%. To Set the Screen Grid Press Display Grid to set the screen grid type. : turn the background grid and coordinate on. : turn the background grid off and coordinate on. : turn the background grid and coordinate off.
RIGOL Chapter 13 Signal Source Chapter 13 Signal Source MSO2000A/DS2000A combines the signal generator and oscilloscope in one instrument by providing a built-in dual-channel, 25MHz signal generator, thus providing convenience for engineers who need to use the signal generator and oscilloscope at the same time. This chapter introduces how to use the built-in signal generator. As the functions and setting methods of source 1 and source 2 are same, this chapter takes Source 1 for illustration.
RIGOL Chapter 13 Signal Source Parameter Setting Method The following two methods are available for setting the basic parameters (such as the frequency, amplitude, start phase and offset) of MSO2000A/DS2000A series oscilloscope. Method 1: Rotate or the navigation knob to set the desired value. Method 2: Press down and the numeric keyboard is displayed, as is shown below. Rotate the knob to select the desired value and unit and then press down the knob to input the value and unit.
Chapter 13 Signal Source RIGOL To Output Basic Waveform To Output Sine Press Setup1 to open the waveform setting menu. Press Wave and select “Sine”. At this point, you can set the sine parameters. 1. Source Press Source to turn on or off the signal output. 2. Set the frequency Press Freq to set the frequency of the signal. The range is from 100 mHz to 25 MHz. Note: For different waveforms, the frequency ranges are different.
RIGOL Chapter 13 Signal Source 5. Set the start phase Press Start to set the start phase of the signal. The range is from 0°to 360°. For the setting method, please refer to the introduction in “Parameter Setting Method”. 6. Align Phase Pressing Align will re-configure the two channels (Source 1 and Source 2) to output according to the preset frequency and phase. For two signals of which the frequencies are the same or are multiples, this operation can align their phase.
Chapter 13 Signal Source RIGOL To Output Ramp Press Setup1 to open the waveform setting menu. Press Wave and select “Ramp”. At this point, you can set the ramp parameters. For the detailed setting method, refer to the introduction in “To Output Sine”. This section only introduces “Symmetry”. Symmetry It is defined as the percentage that the rising period of the ramp takes up in the whole period.
RIGOL Chapter 13 Signal Source To Output Pulse Press Setup1 to open the waveform setting menu. Press Wave and select “Pulse”. At this point, you can set the pulse parameters. For the detailed setting method, refer to the introduction in “To Output Sine”. This section only introduces “DutyCycle”. Duty Cycle It is defined as the percentage that the high level takes up in the whole pulse period.
Chapter 13 Signal Source RIGOL To Output Noise Press Setup1 to open the waveform setting menu. Press Wave and select “Noise”. At this point, you can output noise signal with specified amplitude. 1. Set the amplitude Press Amp to set the amplitude of the signal. When the impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp. For the setting method, please refer to the introduction in “Parameter Setting Method”. 2.
RIGOL Chapter 13 Signal Source To Output Built-in Waveform MSO2000A/DS2000A series oscilloscope provides 7 kinds of built-in waveforms, including Sinc, ExpRise, ExpFall, ECG, Gauss, Lorentz and Haversine.
RIGOL Chapter 13 Signal Source Figure 13-5 ExpFall Function Waveform Figure 13-6 Analog ECG Figure 13-7 Gauss Function Waveform MSO2000A/DS2000A User’s Guide 13-9
RIGOL Chapter 13 Signal Source Figure 13-8 Lorentz Function Waveform Figure 13-9 Haversine Function Waveform Press Setup1 to open the waveform setting menu. Press Wave and select “Built-in”. At this point, you can set the output signal parameters according to the type of the built-in waveform selected. 1. Select built-in waveform Press Built-in to select any of the 7 built-in waveforms. 2. Set the frequency Press Freq to set the frequency of the signal. The range is from 100 mHz to 1 MHz.
Chapter 13 Signal Source RIGOL 3. Set the amplitude Press Amp to set the amplitude of the signal. When the impedance is set to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 Ω, the range is from 10 mVpp to 2.5 Vpp. For the setting method, please refer to the introduction in “Parameter Setting Method”. 4. Set the DC offset voltage Press Offset to set the DC offset of the signal. When the impedance is HighZ, the range is from (-2.5 V+the current amplitude/2) to (2.
RIGOL 8. Chapter 13 Signal Source Set the impedance Press Impedance to set the output impedance of the signal generator. It can be set to “HighZ” or “50Ω”. To Output Arbitrary Waveform MSO2000A/DS2000A allows users to define their own arbitrary waveform and save it in the internal or external memory. At most 10 arbitrary waveforms (LocalArb0.arb to LocalArb9.arb) can be stored in the internal memory. The user-defined waveform can contain 2 to 16384 data points (namely 2 pts to 16 kpts). 1.
Chapter 13 Signal Source RIGOL For the setting method, please refer to the introduction in “Parameter Setting Method”. 5. Align Phase Pressing Align will re-configure the two channels (Source 1 and Source 2) to output according to the preset frequency and phase. For two signals of which the frequencies are the same or are multiples, this operation can align their phase. Acquire the waveforms of the two channels using the oscilloscope and display the waveforms stably.
RIGOL Chapter 13 Signal Source To Select Waveform Press Select to open the waveform selection menu. 1. 2. To Select Stored Waveform Press Stored to enter the file recall interface. Users can select the arbitray waveform stored in the internal or external memory to recall. Use to select the desired waveform and press Load. To Select Channel Waveform Press Channel Source to select the desired source (CH1, CH2 or MATH). Press Channel Range to select “Screen” or “Cursor”.
RIGOL Chapter 13 Signal Source 2. Interpolation Press Interp to turn on or off the interpolation mode between the points of the waveform. ON: the waveform editor connects two points using a straight line. OFF: the waveform editor will keep a constant voltage level between the two points and create a ladder-like waveform. 3. Zoom Press Zoom to turn on or off the zoom function. ON: the waveform editing window only displays the current point.
RIGOL Chapter 13 Signal Source point. This setting is limited by the times of the previous point and following point. The time of point 1 is fixed at 0 s. For the setting method, please refer to the introduction in “Parameter Setting Method”. 7. Insert Press Insert to insert a new waveform point between the current point and the next point. 8. Delete Press Delete to delete the current point from the waveform and connect the remaining points. Note: Point 1 cannot be deleted. 9.
Chapter 13 Signal Source RIGOL two points and create a ladder-like waveform. 2. Zoom Press Zoom to turn on or off the zoom function. ON: the waveform editing window only displays the current point. OFF: the waveform editing window displays all the initial points. 3. Current Point Press Current and use or the navigation knob to select the point to be edited. The range is from 1 to the initial points. For the setting method, please refer to the introduction in “Parameter Setting Method”. 4.
RIGOL Chapter 13 Signal Source “Store and Recall” to save the current waveform file in “.arb” format in the internal memory (you can overwrite the original file or save the current waveform again). You can select the arbitrary waveforms in the internal or external memory for output. For the details, refer to the introduction in “To Select Waveform”.
RIGOL Chapter 13 Signal Source Modulation The built-in signal generator of MSO2000A/DS2000A series oscilloscope supports amplitude modulation (AM) and frequency modulation (FM). The modulated waveform consists of the carrier waveform and modulating waveform. The carrier waveform is the waveform output from the signal generator and the modulating signal can be the built-in sine, square, ramp waveform or noise signal. Press Modulation to open the modulation setting menu.
RIGOL Chapter 13 Signal Source 1. Select the Carrier Waveform Press Setup1 to enter the waveform setting interface. Press Wave to select the desired carrier. Note: When “Pulse”, “DC” or “Noise” is selected, the modulation menu will be hidden automatically. 2. Set the Carrier Parameters After selecting the desired carrier, you can refer to the introduction in “Parameter Setting Method” to set the carrier parameters (frequency, amplitude, etc). 3.
RIGOL Chapter 13 Signal Source FM FM (Frequency Modulation), namely the frequency of the carrier waveform changes with that of the modulating waveform, as shown in the figure below. Modulating Waveform Carrier Waveform Modulated Waveform Figure 13-11 Frequency Modulation 1. Select the Carrier Waveform Press Setup1 to enter the waveform setting interface. Press Wave to select the desired carrier. Note: When “Pulse”, “DC” or “Noise” is selected, the modulation menu will be hidden automatically. 2.
RIGOL Chapter 13 Signal Source 1 Hz to 50 kHz. 5. Set the Modulation Deviation Press Deviation to set the deviation of the frequency of the modulating waveform relative to the carrier frequency. The range is from 0 Hz to the current carrier frequency and the sum of the frequency deviation and carrier frequency cannot be greater than the current carrier frequency upper limit. For the setting method, please refer to the introduction in “Parameter Setting Method”.
Chapter 14 Store and Recall RIGOL Chapter 14 Store and Recall Users can save the current settings, waveforms, screen image and parameters of the oscilloscope in internal memory or external USB mass storage device (such as USB storage device) in various file formats and recall the stored settings or waveforms when needed.
RIGOL Chapter 14 Store and Recall Storage System Press Storage to enter the store and recall setting interface. This oscilloscope provides a USB Host interface at the front panel to connect USB storage device for external storage. The USB storage device connected is marked as “Disk D”. The internal memory (Local Disk) of the instrument can store 10 setting files, 10 reference waveform files and 10 mask files of the Pass/Fail test. Below is the disk selecting interface.
Chapter 14 Store and Recall RIGOL Storage Type Press Storage Storage to select the desired storage type. The default is “Picture”. The storage and recall descriptions of each type are as follows. 1. Traces Save the waveform data in external memory in “*.trc” format. The data of all the channels turned on can be saved in the same file. At recall, the data will be displayed on the screen directly. 2. Waveforms Save the waveform data in external memory in “*.wfm” format.
RIGOL Chapter 14 Store and Recall Tip After a USB storage device is connected, press at the front panel to quickly save the current screen image under the root directory of the USB storage device. 5. CSV Save the waveform data displayed on the screen or of the specified channels in external memory in a single “*.csv” file. You can specify the file name and the saving directory and save the corresponding parameter file (*.txt) under the same directory using the same file name.
RIGOL Chapter 14 Store and Recall Internal Storage and Recall Internal storage and recall support “Setups” in Storage. In the following part, the storage and recall method and procedures are introduced. 1. Save the specified oscilloscope setting in internal memory. 1) Connect the signal to the oscilloscope and obtain stable display. 2) Press Storage Storage to select “Setups” and press Save to turn on the disk management interface.
RIGOL 2. Chapter 14 Store and Recall Load the specified type of file in internal memory. 1) Press Storage Storage to select “Setups” and then press Load to turn on the disk management interface. Use to select “Local Disk” and then press down to open the local disk, as shown in Figure 14-3. 2) Use to select the desired file to load and press Load to load the file selected.
Chapter 14 Store and Recall RIGOL External Storage and Recall Before using external storage and recall, make sure that the USB storage device is connected correctly. External storage supports all the types of files in Storage but in recall, “Picture” and “CSV” are not supported. 1. Save the specified type of file in the external USB storage device (take CSV as an example). 1) Connect the signal to the oscilloscope and obtain stable display.
RIGOL Chapter 14 Store and Recall Figure 14-5 External Disk (To Create a New File) 5) Press OK to execute the saving operation. 2. Load the specified type of file in the external USB storage device (take trace storage as an example). 1) Press Storage Storage to select “Traces” and then press Load to turn on the disk management interface. Use to select “Disk D” and press down to open the USB storage device, as shown in Figure 14-6.
Chapter 14 Store and Recall RIGOL Disk Management Press Storage Disk.Manage to turn on the disk management interface as shown in Figure 14-1 and use to select the desired disk. The disk currently selected is displayed in green and press down to open the disk selected.
RIGOL Chapter 14 Store and Recall To Create a New File or Folder This operation is only valid in external storage. Before using external disk, make sure that the USB storage device is connected correctly. First, press Storage Disk.Manage and use to select and open the external disk (“Disk D”). Then, select the desired file type and select the desired directory under which to create a new file or folder. The default is the root directory of the USB storage device.
RIGOL Chapter 14 Store and Recall English Input Method For example, create a file or folder with the name “Filename”. 1. Press Keyboard. 1) Use to select English input method “En” and uppercase input state “aA”. 2) Use to input the letter “F”. If the input is wrong, press Delete to delete the character input. 3) Use to select lowercase input state “Aa”. 4) Use to input the remaining letters “ilename”.
RIGOL Chapter 14 Store and Recall Chinese Input Method For example, create a file or folder with the name “文件名”. 1. Press Keyboard. 1) Use to select Chinese input method “中”. Note: Chinese is added in the menu items at the right of the screen. 2) Use to input the pinyin “wen”. If the input is wrong, press Delete to delete the pinyin input. After “wen” is input, a series of Chinese characters appear in the “Chinese Chraracter Selecting Area”. 3) Press Chinese and use to select and input “文”.
Chapter 14 Store and Recall RIGOL To Delete a File or Folder Folder operation is valid only in external storage. Before using the external disk, make sure that the USB storage device is connected correctly. 1. Delete a file in internal memory. 1) Press Storage Disk.Manage and use to select and open the local disk (“Local Disk”). 2) Press File Type to select the desired type of file to be deleted (Note: The file types of internal storage include “*.stp”, “*.ref”, “*.pf” and “*.arb”).
RIGOL Chapter 14 Store and Recall To Rename a File or Folder Rename operation is valid only in external storage. Before using the external disk, make sure that the USB storage device is connected correctly. Press Storage Disk.Manage and use to select and open the external disk (“Disk D”). Use to select the desired file or folder to rename and then press Rename to turn on the rename interface. For specific operations, please refer to the descriptions in “To Create a New File or Folder”.
RIGOL Chapter 14 Store and Recall Factory Press Storage Default to return the oscilloscope to its factory state (refer to the table below).
RIGOL Chapter 14 Store and Recall Input 1 MΩ Channel Invert OFF Amplitude Scale Coarse Channel Unit V Acquisition Setting (Acquire) Acquisition Mode Normal Sampling Mode Real Time Memory Depth Auto Anti-Aliasing OFF Trigger Setting (TRIGGER) Trigger Level 0 Trigger Type Edge Source CH1 Slope Rising Edge Trigger Mode Auto Trigger Coupling DC Trigger Holdoff 100 ns Noise Reject OFF Display Setting (Display) Display Type Vectors Persistence Time Min Waveform Intensity 50%
RIGOL Chapter 14 Store and Recall Frequency 1.00 kHz Amplitude 5.00 V Offset Start Phase 0.00 V 0.0° Modulation OFF Impedance HighZ Cursor Setting (Cursor) Mode OFF Manual Display Mode X-Y Source CH1 Screen Region Main Time Unit s A->X -4.000 μs B->X 4.000 μs A->Y 0.000 V B->Y -4.000 V Track Cursor A CH1 Cursor B CH1 A->X -4.000 μs B->X 4.
RIGOL Chapter 14 Store and Recall Y Mask 0.
RIGOL Chapter 14 Store and Recall Invert OFF Vertical Scale 2V A-B Source A CH1 Source B CH1 Invert OFF Vertical Scale A×B 2V Source A CH1 Source B CH1 Vertical Invert OFF Scale A÷B 2U Source A CH1 Source B CH1 Invert OFF Vertical Scale 2U FFT Source CH1 Window Function Rectangle Display Split Scale dB Vertical Scale 20 dB Horizontal Scale 1.25 MHz/div Sample Rate 50 MSa/s Logic Operation Log.
RIGOL Vertical Scale Chapter 14 Store and Recall 2V LA[2] Channel/Group D0 On/Off D7-D0: OFF; D15-D8: OFF Waveform Size M Reorder D0-D15 Threshold Type TTL Digital Bus BUS1 BUS Status OFF Channel Selection D7-D0 Endian Normal Reference Clock None Slope Rising Edge Display Type Data Format Hex Jitter Rejection ON Jitter Time 5.
RIGOL Chapter 14 Store and Recall Data Bits 8 Stop Bit 1 bit Even-Odd Check None Packet OFF Packet End 00 (NULL) I2C SCLK CH1 SDA CH2 Include R/W Close SPI SCLK Channel CH1 SCLK Slope Rising Edge SDA Channel CH2 SDA Polarity 1 for high level CAN Source CH1 Signal Type Rx Baud 500 kb/s Sample Point 50.
Chapter 15 System Function Setting RIGOL Chapter 15 System Function Setting The contents of this chapter: Remote Interface Configuration System-related MSO2000A/DS2000A User’s Guide 15-1
RIGOL Chapter 15 System Function Setting Remote Interface Configuration MSO2000A/DS2000A can communicate with PC via LAN, USB or GPIB (with the USB-GPIB interface converter provided by RIGOL) bus. Please refer to the introduction below to configure the corresponding interface before using the remote interfaces. LAN Setting Press Utility IO Setting LAN Set to turn on the LAN setting interface. You can view the network connection status and configure the network parameters.
Chapter 15 System Function Setting RIGOL IP Configuration Type The configuration type of the IP address can be DHCP, auto IP or static IP. In different IP configuration type, the configuration mode of the network parameters (such as the IP address) is different. DHCP Press Config Mode and use to select “DHCP”. Then press down to select this type. When DHPC type is valid, the DHCP server in the current network will assign the network parameters (such as the IP address) for the oscilloscope.
RIGOL 1. Chapter 15 System Function Setting Set the IP Address The format of IP address is nnn.nnn.nnn.nnn; wherein, the range of the first nnn is from 0 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255. You are recommended to ask your network administrator for an IP address available. Press IP Address and use to input the desired IP address.
Chapter 15 System Function Setting RIGOL Set the Domain Name Server You can set this paramter in Auto IP and Static IP mode. The address format of the domain name server is nnn.nnn.nnn.nnn; wherein, the range of the first nnn is from 0 to 223 (except 127) and the ranges of the other three nnn are from 0 to 255. You are recommended to ask your network administrator for an address available. press DNS and use to input the desired address.
RIGOL Chapter 15 System Function Setting USB Device This oscilloscope can communicate with PC or PictBridge printer via the USB Device interface at the rear panel. You need to set the oscilloscope to make it match with different device types. Press Utility IO Setting USB Device and use device type. to select the desired Computer: in this type, the oscilloscope can communicate with the PC. PictBridge: in this type, the oscilloscope can communicate with the PictBridge printer.
Chapter 15 System Function Setting RIGOL System-related Sound When the sound is enabled, you can hear the sound of the beeper when you press a function key or a menu softkey or when the prompt message pops up. Press Utility Sound to select (on) or (off). The default is off. When the sound is turned on, will be displayed at the lower right corner of the screen. Language This oscilloscope supports multiple language menus, Chinese/English help and prompt messages.
RIGOL Chapter 15 System Function Setting System Information Press Utility System System Info to view the version information of your oscilloscope. The system information contains the following contents as shown in the figure below. Figure 15-2 System Information Interface Power-off Recall You can set the system configuration to be recalled when the oscilloscope is powered on again after power-off. Press Utility System Power On to select “Last” or “Default”.
Chapter 15 System Function Setting RIGOL System Time The system time is displayed at the lower right corner of the screen in “hh:mm (hour:minute)” format. When printing or storing a waveform, the file output will contain this time information. Press Utility System System Time System Time to turn on the setting interface as follows. Figure 15-3 System Time Setting Interface In the figure, the item in green (such as 2013) is the item that can be modified currently.
RIGOL Chapter 15 System Function Setting Screen When the oscilloscope enters the idle state and holds for a certain period of time, the screen saver program will be enabled. Press Utility System Screen to open the screen saver setting menu. “Default” means using the icon RIGOL as the screen saver icon. Press Time to select the screen saver time. When “OFF” is selected, the screen saver program is disabled.
Chapter 15 System Function Setting RIGOL Self-calibration The self-calibration program can quickly make the oscilloscope reach the best working state to get the most precise measurement values. You can perform self-calibration at any time especially when the change of the environment temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been warmed up or operated for more than 30 minutes before the self-calibration.
RIGOL Chapter 15 System Function Setting Print Setting PictBridge is a new print standard. If your oscilloscope and the printer both comply with the PictBridge standard, you can connect the oscilloscope to the printer using a USB cable to print the screen image directly. Devices complies with the PictBridge standard always have the icon as shown in the figure on the right side. MSO2000A/DS2000A digital oscilloscope supports PictBridge printer.
Chapter 15 System Function Setting RIGOL 5. Print Range Press this key to set the print range to “Screen” or “Wave” and the default is “Screen”. Screen: print the whole screen image. Wave: only print the waveform area. 6. Palette Press this key to set the print color to “Gray Scale” or “Color” and the default is “Color”. 7. Paper Size Press this key to open the size list and rotate the multifunction knob to select the desired paper size.
RIGOL Chapter 15 System Function Setting 11. Copies Press this key to set the number of copies to be printed. Rotate the multifunction knob to select the desired number of copies. The range is from 1 to 999. 12. Invert Press this key to set the print color invert to “ON” or “OFF” and the default is “OFF”. When the invert is turned on, the print effect is as shown in the figure below.
Chapter 15 System Function Setting RIGOL Power Status Connect the oscilloscope to AC power using a power cord and at this point, the oscilloscope is energized. MSO2000A/DS2000A oscilloscope allows users to set the power status of the oscilloscope after it is energized. Press Utility PowerStatus to set the power status to “Default” or “Open”. Default: after the oscilloscope is energized, press the power key at the front panel to start the instrument.
RIGOL Chapter 15 System Function Setting Aux Output Users can set the type of the signal output from the [Trigger Out] connector at the rear panel. Press Utility AuxOutput to select the desired output type. TrigOut After this type is selected, the oscilloscope outputs a signal that can reflect the current capture rate of the oscilloscope at each trigger. Connect this signal to a waveform display device, measure the signal frequency and the measurement result equals the current acquisition rate.
RIGOL Chapter 15 System Function Setting Option Management This oscilloscope provides multiple options to fulfill your measurement requirements. Please contact your RIGOL sales representative or RIGOL technical support to order the corresponding options. You can view the options currently installed on the oscilloscope or activate the newly bought option serial number through this menu. Press Utility Options Installed to view the options currently installed on the oscilloscope.
Chapter 16 Remote Control RIGOL Chapter 16 Remote Control MSO2000A/DS2000A oscilloscope can be controlled remotely mainly through the following two methods. User-defined programming Users can program and control the oscilloscope by using the SCPI (Standard Commands for Programmable Instruments) commands. For more information about the commands and programming, refer to the Programming Guide.
RIGOL Chapter 16 Remote Control Remote Control via USB 1. Connect the device Connect the oscilloscope (USB DEVICE) with your PC using a USB cable. 2. Install the USB driver This oscilloscope is a USB-TMC device. Assuming that your PC has already been installed with Ultra Sigma, after you connect the oscilloscope to the PC and turn both on for the first time (the oscilloscope is automatically configured to USB interface), the New Hardware Wizard as shown in the figure below is displayed on the PC.
RIGOL Chapter 16 Remote Control 3 4 5 6 MSO2000A/DS2000A User’s Guide 16-3
RIGOL Chapter 16 Remote Control 7 3. Search device resource Start up Ultra Sigma and the software will automatically search for the oscilloscope resources currently connected to the PC. You can also click to search the resources. 4.
Chapter 16 Remote Control RIGOL and the model number and USB interface information of the instrument will also be displayed as shown in the figure below. Figure 16-1 View the Device Resource under Control via USB 5. Communication test Right click the resource name “MSO2302A (USB0::0x1AB1::0x04B0::DS2A0000000000::INSTR)” to select “SCPI Panel Control” to turn on the remote command control panel (as shown in the figure below) through which you can send commands and read data.
RIGOL Chapter 16 Remote Control Remote Control via LAN 1. Connect the device Connect the oscilloscope to your LAN using a network cable. 2. Configure network parameters Configure the network parameters of the oscilloscope according to the description in “LAN Setting”. 3. Search device resource 16-6 1) Start the Ultra Sigma, click displayed. 2) to search for the instrument resource connected to the local Click network.
Chapter 16 Remote Control 4. RIGOL View device resource The resources found will appear under the “RIGOL Online Resource” directory as shown in the figure below. Figure 16-3 View the Device Resource under Control via LAN 5. Communication test Right click the resource name “MSO2302A(TCPIP::172.16.3.100::INSTR)” to select “SCPI Panel Control” to turn on the remote command control panel (as shown in the figure below) through which you can send commands and read data.
RIGOL Chapter 16 Remote Control through Ultra Sigma (right-click the resource name and select LXI-Web; or directly input the IP address in the browser). Various important information about the oscilloscope (including the model number, manufacturer, serial number, description, MAC address and IP address) will be displayed on the webpage as shown in the figure below.
RIGOL Chapter 16 Remote Control Remote Control via GPIB 1. Connect the device Use the USB to GPIB interface converter to extend a GPIB interface for the oscilloscope. Then connect the oscilloscope to your PC using a GPIB cable. 2. Install the driver of GPIB card Install the driver of the GPIB card which has been connected to the PC correctly. 3. Set the GPIB address Set the GPIB address of the oscilloscope according to the description in “GPIB Address”. 4.
RIGOL Chapter 16 Remote Control problem. 5. View device resource Click “OK” to return back to the main interface of Ultra Sigma. The resources found will appear under the “RIGOL Online Resource” directory. Figure 16-7 View the Device Resource under Control via GPIB 6. Communication Test Right-click the resource name “MSO2302A (GPIB0::1::INSTR)” to select “SCPI Panel Control” to turn on the remote command control panel through which you can send commands and read data as shown in the figure below.
Chapter 17 Troubleshooting RIGOL Chapter 17 Troubleshooting The commonly encountered failures and their solutions are listed below. When you encounter those problems, please solve them following the corresponding steps. If the problem remains still, please contact RIGOL and provide your device information (Utility System System Info). 1. The screen is still dark after pressing the power key: (1) Check whether the power switch is really on. (2) Check whether the power is correctly connected.
RIGOL Chapter 17 Troubleshooting (4) Change the trigger holdoff setting. 5. No display after pressing RUN/STOP: Check whether the trigger mode in the trigger control area (TRIGGER) at the front panel is “Normal” or “Single” and whether the trigger level exceeds the waveform range. If yes, set the trigger level to the middle or set MODE to “Auto”. Note: Using AUTO could automatically finish the above setting. 6. The display of waveform is ladder-like: (1) The horizontal time base might be too low.
Chapter 18 Specifications RIGOL Chapter 18 Specifications All the specifications are guaranteed except the parameters marked with “Typical” and the oscilloscope needs to operate for more than 30 minutes under the specified operation temperature.
RIGOL Maximum Input Voltage (1MΩ) Chapter 18 Specifications Maximum Input Voltage of the Analog Channel CAT I 300 Vrms, CAT II 100 Vrms, Transient Overvoltage 1000 Vpk with RP2200 10:1 probe: CAT II 300 Vrms with RP3300 10:1 probe: CAT II 300 Vrms with RP3500 10:1 probe: CAT II 300 Vrms with RP5600 10:1 probe: CAT II 300 Vrms Digital channel: CAT I 40Vrms, Transient Overvoltage 800 Vpk Horizontal Timebase Scale MSO/DS2302A/2302A-S: 1 ns/div to 1.000 ks/div MSO/DS2202A/2202A-S: 2.000 ns/div to 1.
RIGOL Chapter 18 Specifications MSO/DS2072A/2072A-S: DC to 70 MHz Single-shot Bandwidth MSO/DS2302A/2302A-S: DC MSO/DS2202A/2202A-S: DC MSO/DS2102A/2102A-S: DC MSO/DS2072A/2072A-S: DC to to to to 300 MHz 200 MHz 100 MHz 70 MHz Vertical Resolution Analog channe: 8 bit Digital channel: 1 bit Vertical Scale When the input impedance is 50Ω: 500 μV/div to 1 V/div When the input impedance is 1MΩ: 500 μV/div to 10 V/div Offset Range When the input impedance is 50Ω: 500 μV /div to 50 mV/div: ± 2 V 51 mV/
RIGOL Chapter 18 Specifications Vertical (Digital Channel) Threshold Adjustable threshold of 8 channels per group Threshold Selection TTL (700 mV) 5.0 V CMOS (+2.5 V), 3.3 V CMOS (+1.65 V) 2.5 V CMOS (+1.25 V), 1.8 V CMOS (+900 mV) ECL (-650 mV) PECL (+1.85 V) LVDS (+600 mV) 0V User Threshold Range ±20.
Chapter 18 Specifications Trigger Sensitivity RIGOL 1 div (below 10 mV or noise rejection is enabled) 0.
RIGOL Chapter 18 Specifications Setting Delay Trigger (Option) Edge Type Rising, Falling Delay Type >, <, <>, >< Delay Time 2 ns to 4 s TimeOut Trigger (Option) Edge Type Rising, Falling, Rising/Falling Timeout time 16 ns to 4 s Duration Trigger (Option) Pattern H, L, X Trigger Condition >, <, <> Duration Time 2 ns to 4 s Setup/Hold Trigger Edge Type Rising, Falling Data Type H, L Setup Time 2 ns to 1 s Hold Time 2 ns to 1 s RS232/UART Trigger Polarity Normal, Invert Trigger Cond
RIGOL Chapter 18 Specifications Data Bits 4 bit to 32 bit Data Line Setting H, L, X CAN Trigger (Option) Signal Type Rx, Tx, CAN_H, CAN_L, Differential Trigger Condition SOF, EOF, FrameType, FrameError Signal Rate 10 kbps, 20 kbps, 33.3 kbps, 50 kbps, 62.5 kbps, 83.
RIGOL Chapter 18 Specifications Frequency, Period, Positive Pulse Width, Negative Pulse Width, Positive Duty Cycle, Negative Duty Cycle, Delay 1 2 , Delay 1 2 , Delay 1 2 , Delay 1 2 , Phase 1 2 , Phase 1 2 , Phase 1 2 , Phase 1 2 Number of Measurements Display 5 measurements at the same time.
RIGOL Chapter 18 Specifications Display Color 160,000 Color (TFT) Persistence Time Min, 50ms, 100ms, 200ms, 500ms, 1 s, 2 s, 5 s, 10 s, 20 s, Infinite Display Type Dots, Vectors Real-time Clock Time and Date (user adjustable) Signal Source (MSO2000A-S/DS2000A-S) Channels 2 Sample Rate 200 MSa/s Vertical Resolution 14 bits Max.
RIGOL Chapter 18 Specifications Min.
RIGOL Chapter 18 Specifications General Specifications Probe Compensation Output Output Voltage[1] Frequency [1] About 3 V, peak-peak 1 kHz Power Power Voltage 100 V to 240 V, 45 Hz to 440 Hz Power Maximum 50 W Fuse 2 A, T Degree, 250 V Environment Temperature Range Operating: 0 ℃ to +50 ℃ Non-operating: -40 ℃ to +70 ℃ Cooling Method Humidity Range Fan cooling 0℃ to +30℃: ≤95% Relative Humidity +30℃ to +40℃: ≤75% Relative Humidity +40℃ to +50℃: ≤45% Relative Humidity Altitude Operating: unde
RIGOL Chapter 19 Appendix Chapter 19 Appendix Appendix A: Accessories and Options Order Description Model Number DS2302A 300MHz 2-analog channel oscilloscope DS2302A DS2202A 200MHz 2-analog channel oscilloscope DS2202A DS2102A 100MHz 2-analog channel oscilloscope DS2102A DS2072A 70MHz 2-analog channel oscilloscope DS2072A DS2302A-S 300MHz DS2202A-S 200MHz DS2102A-S 100MHz DS2072A-S 70MHz MSO2302A 300 MHz MSO2202A 200 MHz MSO2102A 100 MHz MSO2072A 70 MHz 2-analog channel
RIGOL Chapter 19 Appendix 2-analog channel + 16-digital MSO2072A-S 70MHz channel mixed signal oscilloscope MSO2072A-S + 2-channel 25MHz signal source Power Cord conforming to the standard of the - country Standard Accessories Accessories CB-USBA-USB USB Cable B-FF-150 2 Passive Probes (350 MHz) RP3300A A logic analyzer probe RPL2316 Quick Guide - Resource CD (include User’s Guide and application software) Optional Passive Probe (500 MHz) RP3500A Accessories Rack Mount Kit RM-DS-2000A
Chapter 19 Appendix RIGOL Appendix B: Warranty RIGOL warrants that its products mainframe and accessories will be free from defects in materials and workmanship within the warranty period. If a product is proven to be defective within the respective period, RIGOL guarantees the free replacement or repair of products which are approved defective. To get repair service, please contact with your nearest RIGOL sales and service office.
Index RIGOL Index - Duty ...................................... 6-14 - Width .................................... 6-14 + Duty..................................... 6-14 + Width ................................... 6-14 Acquisition Mode ........................ 4-2 Amplitude Modulation.............. 13-19 AND .......................................... 6-7 Antialiasing ................................ 4-9 Area ........................................ 6-18 Auto IP ....................................
RIGOL Nth Edge Trigger ...................... 5-18 Operator .................................. 6-11 OR ............................................ 6-7 Overshoot ................................ 6-18 Parallel Decoding ........................ 8-2 pass/fail test ............................ 10-1 Pattern Setting ................. 5-26, 5-33 Pattern Trigger ......................... 5-26 Peak Detect ............................... 4-4 Period ..................................... 6-14 Period Area .............
Index Vtop ........................................ 6-17 Waveform Analysis .................... 11-8 Waveform Confusion ................... 4-6 Waveform Distortion ................... 4-6 Waveform Invert......................... 2-5 Waveform Leakage ..................... 4-6 Waveform Playback ................... 11-6 Waveform Record ..................... 11-2 MSO2000A/DS2000A User’s Guide RIGOL Waveforms ............................... 14-3 Window Function........................ 6-4 Windows trigger ...