User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
RIGOL Chapter 3 To Set the Horizontal System
3-4 MSO2000A/DS2000A User’s Guide
Time Base Mode
Press MENU in the horizontal control area (HORIZONTAL) at the front panel and
then press Time Base to select the time base mode of the oscilloscope and the
default is Y-T.
Y-T Mode
In this mode, the Y axis represents voltage and the X axis represents time.
Note: Only when this mode is enabled can “Delayed Sweep” be turned on. In this
mode, when the horizontal time base is greater than or equal to 200 ms, the
instrument enters slow sweep mode. For the details, please refer to the introduction
of slow sweep in “Roll Mode”.
X-Y Mode
In this mode, the oscilloscope changes the two channels from voltage-time display
mode to voltage-voltage display mode. Wherein, the X axis and Y axis tracks the
voltages of CH1 and CH2 respectively. The phase deviation between two signals with
the same frequency can be easily measured via Lissajous method. The figure below
shows the measurement schematic diagram of the phase deviation.
Figure 3-2 Measurement Schematic Diagram of Phase Deviation
II I
III IV
A
B
C
D
The signal must be
centered horizontally
II I
III IV