User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
RIGOL Chapter 17 Troubleshooting
MSO2000A/DS2000A User’s Guide
17-2
(4) Change the trigger coupling type: if edge trigger is currently selected, press
MENU (in the trigger control area (TRIGGER) at the front panel)
Setting Coupling to select “HF Reject” or “LF Reject” to filter out the
high-frequency or low-frequency noise that disturbs the trigger.
(5) Change the trigger holdoff setting.
5. No display after pressing RUN/STOP:
Check whether the trigger mode in the trigger control area (TRIGGER) at the
front panel is “Normal” or “Single” and whether the trigger level exceeds the
waveform range. If yes, set the trigger level to the middle or set MODE to
“Auto”.
Note: Using AUTO could automatically finish the above setting.
6. The display of waveform is ladder-like:
(1) The horizontal time base might be too low. Increase the horizontal time
base properly to improve the display.
(2) If the display type is “Vectors”, the lines between the sample points may
cause ladder-like display. Press Display Type to set the display type to
“Dots” to solve the problem.
7. Digital waveform is not stable:
(1) Check whether the signal input from the selected trigger source is
synchronized with the digital signal.
(2) Adjust the oscilloscope to select the appropriate trigger mode, time base
and other general settings.
(3) If the display is still not stable, please check the electrical connections and
parameter settings again.
(4) Try to use other probe (such as analog probe) to verify the signal status of
the test point.
8. Fail to connect PC or PictBridge printer through USB:
(1) Press Utility IO Setting USB Device to check whether the current
setting matches the device currently connected.
(2) Check whether the USB cable is correctly connected to the oscilloscope.
(3) Check whether the USB cable is in good condition. If needed, restart the
oscilloscope.