User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
RIGOL Chapter 13 Signal Source
13-12 MSO2000A/DS2000A User’s Guide
8. Set the impedance
Press Impedance to set the output impedance of the signal generator. It can
be set to “HighZ” or “50Ω”.
To Output Arbitrary Waveform
MSO2000A/DS2000A allows users to define their own arbitrary waveform and save it
in the internal or external memory. At most 10 arbitrary waveforms (LocalArb0.arb to
LocalArb9.arb) can be stored in the internal memory. The user-defined waveform
can contain 1 to 16384 data points (namely 1 pts to 16 kpts).
1. Set the frequency
Press Freq to set the frequency of the signal. The range is from 100 mHz to 10
MHz. For the setting method, please refer to the introduction in “
Parameter
Setting Method”.
2. Set the amplitude
Press Amp to set the amplitude of the signal. When the impedance is set to
HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 Ω,
the range is from 10 mVpp to 2.5 Vpp. For the setting method, please refer to
the introduction in “
Parameter Setting Method”.
3. Set the DC offset voltage
Press Offset to set the DC offset of the signal. When the impedance is HighZ,
the range is from (-2.5 V + the current amplitude/2) to (2.5 V - the current
amplitude/2); when the impedance is set to 50 Ω, the range is from (-1.25 V +
the current amplitude/2) to (1.25 V - the current amplitude/2). For the setting
method, please refer to the introduction in “
Parameter Setting Method”.
Note: The offset value only contains three effective digits. For example, when
the amplitude is set to 30 mVpp (the impedance is 50 Ω), the offset range
calculated is from -1.235 V to 1.235 V, while the actual offset range is -1.23 V to
1.23 V.
4. Set the start phase
Press Start to set the start phase of the signal. The range is from 0° to 360°.