User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
Chapter 11 Waveform Record RIGOL
MSO2000A/DS2000A User’s Guide 11-11
(refer to the setting method of the current frame mentioned above) and the
default is the first frame.
End Frame: press this key to set the end frame of waveform analysis (refer
to the setting method of the current frame mentioned above) and the
default is the last frame.
Threshold: press this key to set the threshold of waveform analysis and the
range is from 1% to 99%. The threshold is used to judge whether the frame
is an error frame. A frame is recognize as an error one if the (relative)
difference value between the frame and the templet is equal to or larger
than the threshold currently set.
9. Time Tag
The time tag is used to display the absolute recording time of each frame of the
waveform currently recorded. Press Time Tag to enable or disable the time tag
function. When it is enabled, the time tag information is displayed at the upper
right corner of the screen as shown in the figure below.
T
0
:
display the absolute starting time of the waveform record.
△T: display the recording time deviation of the current waveform relative to the
first frame of waveform.