User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
Chapter 8 Protocol Decoding RIGOL
MSO2000A/DS2000A User’s Guide 8-15
(positive polarity) or (negative polarity).
If the CS channel is set to CH1 or CH2, you can press Threshold to set the
threshold of the clock channel. The range of the threshold is related to the
vertical scale and vertical position and is from (-5*vertical scale - vertical
position) to (5*vertical scale - vertical position). If the CS channel is set to
any of D0-D15, Threshold will be hidden automatically.
2. SCLK Setting
Press SCLK to enter the clock line setting interface.
Press Channel to select any channel (CH1, CH2 or any channel of D0-D15)
as the clock channel.
Press Slope to set to sample the SDA on the rising or falling edge of the
SCLK.
If the clock channel is set to CH1 or CH2, press Threshold to set the
threshold of the clock channel. The range of the threshold is decided by the
vertical position and scale, and is from (-5*vertical scale - vertical position)
to (5*vertical scale - vertical position). If the clock channel is set to any
channel of D0-D15, Threshold will be hidden automatically.
3. SDA Setting
Press SDA to enter the SDA data line setting interface.
Press Channel to select any channel (CH1, CH2 or any channel of D0-D15)
as the SDA data channel.
Press Polarity to set the polarity of the SDA data line to (high) or
(low).
If the data channel is set to CH1 or CH2, press Threshold to set the
threshold of the data channel. The range of the threshold is decided by the
vertical position and scale, and is from (-5*vertical scale - vertical position)
to (5*vertical scale - vertical position). If the data channel is set to any
channel of D0-D15, Threshold will be hidden automatically.
4. Data Bits Setting
Press Data Bits to set the number of bits of each frame of data. The range
available is from 4 to 32. The default is 8.
5. Endian Setting
Press Endian to select “LSB” or “MSB” and the default is “MSB”.