User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
Chapter 6 MATH and Measurements RIGOL
MSO2000A/DS2000A User’s Guide 6-39
X-Y Mode
X-Y mode is only available when the main time base is “X-Y” mode. In this mode, you
can adjust the cursor to measure X and Y values of the crossing points of the two
pairs of cursors.
Press MENU at the horizontal control are (HORIZONTAL) at the front panel and then
press Time Base to select “X-Y”. Press Cursor Mode “X-Y” to enable the X-Y
mode cursor measurement function. You can also press MENU or Mode
continuously to switch to “X-Y”. The measurement results are displayed at the upper
left corner of the screen in the following mode.
CurAX: the X value at cursor A.
CurBX: the X value at cursor B.
CurAY: the Y value at cursor A.
CurBY: the Y value at cursor B.
You can adjust the positions of the cursors by the following instructions.
Press Type to select “X”. Then, press CursorA and CursorB and use to adjust
the horizontal positions of cursor A and cursor B respectively. Press CursorAB and
use
to adjust the horizontal position of cursor A and B at the same time. During
the adjustment, the measurement result will change accordingly. The adjustable
range is limited within the screen.
Press Type to select “Y”. Then, press CursorA and CursorB and use to adjust
the vertical positions of cursor A and cursor B respectively. Press CursorAB and use
to adjust the vertical positions of cursor A and cursor B at the same time. During
the adjustment, the measurement result will change accordingly. The adjustable
range is limited within the screen.