User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
RIGOL Chapter 6 MATH and Measurements
6-38 MSO2000A/DS2000A User’s Guide
Auto Mode
In this mode, one or more cursors will appear. You can use auto cursor measurement
to measure any of the 29 waveform parameters. Before using this mode, you need to
at least enable one auto measurement parameter and the number of cursors will
change with the measurement parameter enabled.
Press Cursor Mode “Auto” and the number of cursors displayed on the screen
is determined by the measurement parameter enabled (different measurement
parameter needs different number of cursors).
Note: No cursor will be displayed if no auto cursor measurement parameter is
enabled or the measurement source has no input. When the waveform is expanded
or compressed horizontally, the cursor will move accordingly.
If multiple measurement parameters are turned on later, you can use Meas.Para. to
switch among at most five measurement parameters that are turned on last and the
measurement parameter currently selected is high-lighted at the bottom of the
screen.
The following figure shows the auto measurement of the period of CH1.