User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
Chapter 6 MATH and Measurements RIGOL
MSO2000A/DS2000A User’s Guide 6-33
°: in the measurement results, A->X, B->X and △X are expressed in degree.
At this point, A->X, B->X and △X will change to “0°”, “360°” and “360°”
respectively when you press SetCursor, no matter where cursor A and
cursor B are currently located. At the same time, two cursor lines
(unmovable) appear on the screen as the reference positions.
%: in the measurement results, A->X, B->X and △X are expressed in
percentage.
At this point, A->X, B->X and △X will change to “0%”, “100%” and “100%”
respectively when you press SetCursor, no matter where cursor A and B
are currently located. At the same time, two cursor lines (unmovable)
appear on the screen as the reference positions.
Note: When the display mode is set to “Y”, the Time Unit menu will be hidden
automatically and cannot be set.
When the display mode is “Y” or “X-Y”, press Vertical Unit to select “Source
Unit” or “%”.
Source Unit: when this unit is selected, in the measurement results, the
units of A->Y, B->Y and △Y will be automatically set to the unit of the
current source.
%: when this unit is selected, in the measurement results, A->Y, B->Y and
△X are expressed in percentage. At this point, A->Y, B->Y and △X will
change to “0%”, “100%” and “100%” respectively when you press
SetCursor, no matter where cursor A and B are currently located. At the
same time, two cursor lines (unmovable) appear on the screen as the
reference positions.
Note: When the display mode is set to “X” or the measurement source is set to
LA, the Vertical Unit menu will be hidden automatically and cannot be set.
When the measurement source is “LA”
When the measurement source is “LA”, DisplayMode is set to “X” automatically
and cannot be changed; you can only set the X axis unit. Press Time Unit
softkey to select “s”, “Hz”, “°” or “%”.
s: in the measurement results, CurA, CurB and △X denote time.
Hz: in the measurement results, CurA, CurB and △X denote frequency.
°: in the measurement results, CurA, CurB and △X are expressed in degree.
For the description and setting method of phase reference position, please
refer to the related introduction above.