User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
Chapter 6 MATH and Measurements RIGOL
MSO2000A/DS2000A User’s Guide 6-29
Cursor Measurement
Before making cursor measurements, connect the signal to the oscilloscope and
acquire stable display. Cursor measurement can measure the X axis values (usually
Time) and Y axis values (usually Voltage) of the selected waveform. All the
parameters supported by the “Auto Measurement” function can also be measured
through cursors.
The cursor measurement function provides the following two kinds of cursors.
X Cursor
X cursor is a vertical solid/dotted line used to perform horizontal adjustment. It
can be used to measure time (s), frequency (Hz), phase (°) and ratio (%).
― Cursor A is a vertical solid line and cursor B is a vertical dotted line.
― In the X-Y cursor mode, X cursor is used to measure the waveform
amplitude of CH1.
Y Cursor
Y cursor is a horizontal solid/dotted line used to perform vertical adjustment. It
can be used to measure amplitude (the unit is the same as that of the source
channel amplitude) and ratio (%).
― Cursor A is a horizontal solid line and cursor B is a horizontal dotted line.
― In the X-Y cursor mode, Y cursor is used to measure the waveform
amplitude of CH2.
Press Cursor (in the function menu at the front panel) Mode; use to select
the desired cursor mode (the default is “OFF”) and then press down the knob. You
can also press Cursor or Mode continuously to switch the current cursor mode. The
X Cursor (Cursor A)
X Cursor (Cursor B)
Y Cursor
(Cursor B)
Y Cursor
(Cursor A)