User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
RIGOL Chapter 6 MATH and Measurements
6-24 MSO2000A/DS2000A User’s Guide
Measurement Setting
1. Source Selection
Press Measure Source to select the source channel (CH1, CH2, MATH or
any channel of D0-D15) of the time, voltage or area parameter. The color of the
parameter icons under MENU at the left of the screen will change with the
source selected.
2. Measurement Range
Press Measure Range Range to select “Screen” or “Cursor” for
measurement.
When “Cursor” is selected, two cursor lines appear on the screen. At this point,
press CursorA and CursorB and use to adjust the position of the two
cursor lines respectively so as to determine the measurement range. Or, press
CursorAB and use to adjust the positions of cursor A and cursor B at the
same time.
Note: You can press continuously to switch among CursorA, CursorB and
CursorAB.
Figure 6-6 Cursor Region
3. Delayed Measurement Setting
Specify source A and source B in the measurement items “Dly A
→B ”, “Dly
A
→B ”, “Dly A →B ” and “Dly A →B ”.
Press Measure Setting Type “Delay” and then press SourceA and
SourceB to set the two channel sources (CH1, CH2 or any channel of D0-D15)
of delayed measurement respectively.
Note: When SourceA is set to any analog channel (or digital channel),