User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
Chapter 5 To Trigger the Oscilloscope RIGOL
MSO2000A/DS2000A User’s Guide 5-47
USB Trigger (Option)
Trigger on the SOP, EOP, RC, Suspend and Exit Suspend of the data packet on the
differential USB data cable (D+ and D-). This trigger supports USB Low Speed and
Full Speed. The figure below shows the USB data transmission protocol.
Figure 5-16 USB Trigger
Trigger Type:
Press Type to open the trigger type list. Rotate and press down to select
“USB”. At this point, the trigger setting information as shown in the figure below is
displayed at the upper right corner of the screen. For example,
.
The trigger type is USB trigger; the trigger source is CH1; the trigger level is 0.00 V.
Source Selection:
Press D+ and D- to specify data sources for D+ and D- data cables respectively.
They can be set to CH1, CH2 or any channel of D0-D15. For the details, please refer
to the introduction in “
Trigger Source”. The current trigger source is displayed at
the upper right corner of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Signal Speed:
Press Speed to select “Low Speed” (1.5 Mb/s) or “Full Speed” (12 Mb/s).
Trigger Condition:
Press When to select the desired trigger condition.
SOP: trigger at the sync bit at the start of the data packet (SOP).
EOP: trigger at the end of the SEO portion of the EOP of the data packet.
RC: trigger when SEO is greater than 10 ms.
Suspend: trigger when the idle state of the bus is greater than 3 ms.