User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
RIGOL Chapter 5 To Trigger the Oscilloscope
5-28 MSO2000A/DS2000A User’s Guide
: set the pattern of the channel selected to “H”, namely the voltage level is
higher than the threshold level of the channel.
: set the pattern of the channel selected to “L”, namely the voltage level is
lower than the threshold level of the channel.
: set the pattern of the channel selected to “Don’t Care”, namely this channel is
not used as a part of the pattern. When both the channels in the pattern are set
to “Don’t Care”, the oscilloscope will not trigger.
or : set the pattern of the channel selected to the rising or falling edge.
Note: Only one rising or falling edge can be specified in the pattern. If one edge
item is currently defined and then another edge item is defined in another channel in
the pattern, the former edge item defined will be replaced by X.
Trigger Mode:
Press Sweep and select “Auto”, “Normal” or “Single”. For the details, please refer to
“
Trigger Mode”. The corresponding status light of the current trigger mode turns
on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Note: Noise rejection are grayed out and disabled automatically when the signal
source is set to any channel of D0-D15.
Trigger Level:
The trigger level of each channel needs to be set independently. For example, set the
trigger level of CH1. Press Source and select CH1; use TRIGGER LEVEL to
modify the level. For the details, please refer to the introduction of “
Trigger Level”.