User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
Chapter 5 To Trigger the Oscilloscope RIGOL
MSO2000A/DS2000A User’s Guide 5-23
When the Slope Condition is set to , , , or , the
current trigger level and slew rate will be displayed at the lower left corner of the
screen, as shown in Figure 5-7 (a). The formula of slew rate is:
UpLevel - LowLevel
SlewRate=
Time
When the Slope Condition is set to
or , the current trigger level
and slew rate range will be displayed at the lower left corner of the screen, as shown
in Figure 5-7 (b). The formula of slew rate range is:
UpLevel - LowLevel UpLevel - LowLevel
SlewRate= ~
UpperLimit LowerLimit
(a)
(b)
Figure 5-7 Trigger Level Information
Note: Under the “Slope” trigger menu, you can also press down the trigger level
knob continuously to switch the vertical window.
During the adjustment, two orange trigger level lines and two trigger marks (
and
) appear on the screen and move up and down with the rotation of the knob,
when stopping turning the knob, the trigger level lines and trigger marks disappear
in about 2 s.
Trigger Mode:
Press Sweep and select “Auto”, “Normal” or “Single”. For the details, please refer to
“
Trigger Mode”. The corresponding status light of the current trigger mode turns
on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.