User Manual
Table Of Contents
- Guaranty and Declaration
- Safety Requirement
- MSO2000A/DS2000A Series Overview
- Document Overview
- Chapter 1 Quick Start
- Chapter 2 To Set the Vertical System
- Chapter 3 To Set the Horizontal System
- Chapter 4 To Set the Sample System
- Chapter 5 To Trigger the Oscilloscope
- Trigger Source
- Trigger Mode
- Trigger Coupling
- Trigger Holdoff
- Noise Rejection
- Trigger Type
- Edge Trigger
- Pulse Trigger
- Runt Trigger
- Windows Trigger (Option)
- Nth Edge Trigger (Option)
- Slope Trigger
- Video Trigger (HDTV Option)
- Pattern Trigger
- Delay Trigger (Option)
- TimeOut Trigger (Option)
- Duration Trigger (Option)
- Setup/Hold Trigger
- RS232 Trigger
- I2C Trigger
- SPI Trigger
- USB Trigger (Option)
- CAN Trigger (Option)
- Trigger Output Connector
- Chapter 6 MATH and Measurements
- Chapter 7 Digital Channel
- Chapter 8 Protocol Decoding
- Chapter 9 Reference Waveform
- Chapter 10 Pass/Fail Test
- Chapter 11 Waveform Record
- Chapter 12 Display Control
- Chapter 13 Signal Source
- Chapter 14 Store and Recall
- Chapter 15 System Function Setting
- Chapter 16 Remote Control
- Chapter 17 Troubleshooting
- Chapter 18 Specifications
- Chapter 19 Appendix
- Index
RIGOL Chapter 5 To Trigger the Oscilloscope
5-14 MSO2000A/DS2000A User’s Guide
Runt Trigger
Runt trigger is used to trigger pulses that pass one trigger level but fail to pass the
other one.
For this oscilloscope, positive runt pulses are defined as pulses that pass the lower
limit of the trigger level but fail to pass the upper limit of the trigger level; negative
runt pulses are defined as pulses that pass the upper limit of the trigger level but fail
to pass the lower limit of the trigger level, as shown in the figure below.
Figure 5-4 Runt Trigger
Trigger Type:
Press Type, rotate to select “Runt” and press down . At this point, the trigger
setting information is displayed at the upper right corner of the screen. For example,
. The trigger type is runt trigger; the trigger source is CH1; the
difference between the upper limit of trigger level and the lower limit of trigger level
is 3.06 V.
Source Selection:
Press Source to switch the signal source list between CH1 and CH2. For the details,
please refer to the introduction in “
Trigger Source”. The current trigger source is
displayed at the upper right corner of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Pulse Polarity:
Press Polarity to select the pulse polarity of runt trigger.
: positive polarity. The instrument triggers on the positive runt pulse.
: negative polarity. The instrument triggers on the negative runt pulse.
Upper Limit of the
Trigger Level
Lower Limit of the
Trigger Level
Positive Runt Pulse
Negative Runt Pulse