:MEASure:STATistic:MODE
Syntax :MEASure:STATistic:MODE <mode>
:MEASure:STATistic:MODE?
Description
Set or query the statistic mode.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{DIFFerence|EXTRemum}
EXTRemum
Explanation DIFFerence: select the difference mode. The statistic results contain the current
value, average value, standard deviation, and counts.
EXTRemum: select the extremum mode. The statistic results contain the current
value, average value, minimum, and maximum.
Sending the
:MEASure:STATistic:DISPlay command can enable the statistic function.
When the statistic function is enabled, the oscilloscope makes statistic and displays
the statistic results of at most 5 measurement items that are turned on last.
Return
Format
The query returns DIFF or EXTR.
Example
:MEASure:STATistic:MODE DIFFerence /*Select the difference mode*/
:MEASure:STATistic:MODE? /*The query returns DIFF*/
Related
Commands
:MEASure:STATistic:RESet
:MEASure:STATistic:ITEM
:MEASure:STATistic:RESet
Syntax
:MEASure:STATistic:RESet
Description
Clear the history data and make statistic again.
Related
Commands
:MEASure:STATistic:DISPlay
:MEASure:STATistic:MODE
:MEASure:STATistic:ITEM