User`s guide
Chapter 2 Front Panel Operations RIGOL
M300 User’s Guide
2-53
To Control MC3164/MC3264
MC3164 is a 64-channel single-ended multiplexer (MUX-64 for short). MC3264 is a
64-channel single-ended reed multiplexer (REED-MUX-64 for short) and can provide
relatively faster switching speed. MC3164/MC3264 is divided into two banks of 32
single-ended channels (HI only, the 32 channels has a common LO terminal) each
and cannot be used for 4-wire connection. MC3164/MC3264 is used with the external
terminal block (M3TB64) which has a built-in thermocouple reference junction.
01
...
...
DMM
Source
H
L
32
33
64
Bank Switch
Channel Switches
96
Backplane Switches
Com H
Com L
95
Signals
to be
Measured
Reference
Junction
Terminal Box
: M
3
TB64
H
01
H
64
.
.
.
78
-
Pin
Interface
Com L
Figure 2-20 MC3164 Schematic Diagram
01
...
...
DMM
Source
H
L
32
33
64
Bank Switch
Channel Switches
96
Backplane Switches
Com H
Com L
95
Signals
to be
Measured
Reference
Junction
Terminal Box: M3TB64
H
01
H
64
.
.
.
78-Pin
Interface
Com L
Figure 2-21 MC3264 Schematic Diagram
The control interface of MC3164/MC3264 is as shown in Figure 2-22
. Press Page Up
and Page Dn at the front panel to open the previous and next pages of the control
interface. Use the up/down and left/right direction keys to select the desired channel.
The channel currently selected is highlighted, for example,
. You can perform
various operations on the channel currently selected (such as disconnect and close