User`s guide

Chapter 2 Front Panel Operations RIGOL
M300 Users Guide
2-49
5 Bank Separate/Bank Joined
By default, this module is in “Bank Joined” state (namely the bank switch in
Figure 2-14 is closed and the control interface is as shown in Figure 2-15
). At
this point, this module can only be used for 2-wire connection. Press Page Dn
Bank Sepa to execute the “Bank Separateoperation (namely disconnect
the bank switch in
Figure 2-14 and the control interface is as shown in Figure
2-16). At this point, this module is divided into two independent banks and can
be used for 4-wire connection. When it is used for 4-wire resistance
measurement, the instrument pairs channel n and channel n+10 automatically
(wherein, n is an integer between 1 and 10). Channel n is used to connect the
source terminal of DMM and channel n+10 is used to connect the sense terminal
of DMM.
Figure 2-16 MC3120 Control Interface (Bank Separate)