Channel Copy
Channel copy can copy the configuration of a channel (source channel) to another
channel or multiple channels (target channels) of the same type. After that, the
following settings of the source channel are copied to the target channels
automatically.
1 Measurement Configuration: include measurement function, range, etc.
2 Scaling Setting: on/off status of the scaling function and scaling coefficients.
3 Alarm Setting: on/off status of the alarm function, alarm channel number and
alarm limits.
4 Advanced Configuration: include integration time, AC filter, etc.
Press Channel to enter the channel copy interface. This interface is divided into the
upper and lower areas.
Switch: switch the cursor to select desired area. This menu is available only
when the current scan list has 2 or more channels.
Upper Area: use the channel selected as the source channel. You can select the
target channels.
Lower Area: use the channel selected as the target channel. You can select the
source channel.
Done: copy the configuration of the source channel selected to the target
channel selected.
Return: return to the upper-level menu.
Figure 2-10 Channel Copy
Steps of Channel Copy:
After entering the channel copy interface, the channel selected is fixed as the source
channel and the upper area is selected by default. Use the left/right direction keys to
select the desired target channels, press OK to confirm it (press OK again to cancel