User`s guide

RIGOL Chapter 2 Front Panel Operations
2-28 M300 User’s Guide
Table 2-4 Auto Delay Time
(a) DCV/DCI/TEMP (TC)/SENSOR (DCV/DCI) (all the ranges):
Integration Time
[1]
Channel Delay
>1 PLC
2.0 ms
≤1 PLC
1.0 ms
(b) 2WR/4WR/TEMP (THER/RTD/RTD 4W)/SENSOR (2WR/4WR):
Range
Channel Delay
(Integration Time>1
PLC)
[1]
Range
Channel Delay
(Integration Time≤1
PLC)
200 Ω
2.0 ms
200 Ω
1.0 ms
2 kΩ
2.0 ms
2 kΩ
1.0 ms
20 kΩ
2.0 ms
20 kΩ
1.0 ms
200 kΩ
25 ms
200 kΩ
20 ms
1 MΩ
30 ms
1 MΩ
25 ms
10 MΩ
200 ms
10 MΩ
200 ms
100 MΩ
200 ms
100 MΩ
200 ms
Note
[1]
: for DCV, TEMP, 2WR, 4WR and SENSOR (except FREQ sensor), if the current
integration time is set to User, the instrument calculates the channel delay according to the
condition Integration Time>1 PLC automatically.
(c) ACV/ACI (all the ranges):
AC Filter
Channel Delay
3 Hz (slow)
7.0 s
20 Hz (medium)
1.0 s
200 Hz (fast)
120 ms
(d) FREQ/PERIOD/SENSOR (FREQ):
AC Filter
Channel Delay
3 Hz (slow)
600 ms
20 Hz (medium)
300 ms
200 Hz (fast)
100 ms
Delay
Use the numeric keyboard to enter the desired delay. The range is from 0 to 60
s and the resolution is 1 ms.