User`s guide
RIGOL Chapter 1 Quick Guide
1-18 M300 User’s Guide
isolated inputs to the DMM module. MC3232 is divided into two banks (A and B) of 16
two-wire channels. When making 4-wire resistance measurement, the instrument
automatically pairs channels of A bank and B bank. All channels are
break-before-make. You can close multiple channels on this module only if you have
not configured any channel to be part of the scan list. This module can be connected
with MC3065 (DMM Module, if MC3065 is currently inserted). For the specifications of
this module, please refer to the User’s Guide or Data Sheet of this product.
MC3264 (64-Channel Single-Ended Reed Multiplexer):
The MC3264 is applicable to high-speed scan and high-throughput auto test
applications. All 64 channels can switch HI input only. MC3264 is divided into two
banks (A and B) of 32 single-ended channels each. All channels are
break-before-make. In any case, you cannot close multiple channels on this module.
This module can be connected with MC3065 (DMM Module, if MC3065 is currently
inserted) but cannot be used for 4-wire resistance measurement. For the
specifications of this module, please refer to the User’s Guide or Data Sheet of this
product.
MC3324 (24-Channel Multiplexer):
The MC3324 is a mixed multiplexer with 20 voltage channels and 4 current channels.
All 20 voltage channels switch both HI and LO inputs, thus providing fully isolated
inputs to the DMM module, and is divided into two banks (called A and B) of 10
two-wire channels each. When making 4-wire resistance measurement, the
instrument automatically pairs channels of A bank and B bank. All channels are
break-before-make. You can close multiple channels on this module only if you have
not configured any channel to be part of the scan list.
4 current channels are used in combination with the DMM module to measure DC
current or AC current. All channels are break-before-make. In any case, you cannot
close multiple channels on this module. This module can be connected with MC3065
(DMM Module, if MC3065 is currently inserted). For the specifications of this module,
please refer to the User’s Guide or Data Sheet of this product.
MC3416 (16-Channel Actuator):
The MC3416 can connect signals to the device under test or enable external devices.
Any of the 16 channels can switch to Normally-Open (NO) and Normally-Closed (NC)
states. For the specifications of this module, please refer to the User’s Guide or Data
Sheet of this product.
MC3534 (Multifunction Module):
This module provides 3 kinds of functions with 4 channels for each function. This
module can be used to check status or control external devices (such as solenoid,
power relays and microwave switches). You can also read digital inputs and the