User`s guide
RIGOL Chapter 6 Specifications
6-6 M300 User’s Guide
Frequency and Period Characteristics
Accuracy Specifications: ± (% of reading)
[1] [2]
Function
Range
Frequency Range
24 Hour
[3]
T
CAL
℃
±1
℃
90 Day
T
CAL
℃
±5
℃
1 Year
T
CAL
℃
±5
℃
Temperature Coefficient
0
℃
to (T
CAL
℃
-5
℃
)
(T
CAL
℃
+5
℃
)to 50
℃
Frequency,
Period
200mV to
300V
3 Hz-5 Hz
0.07
0.07
0.07
0.005
5 Hz-10 Hz
0.04
0.04
0.04
0.005
10 Hz-40 Hz
0.02
0.02
0.02
0.001
40 Hz-300 kHz
0.005
0.006
0.007
0.001
300 kHz-1 MHz
0.005
0.006
0.007
0.001
Additional Low Frequency Errors: (% of reading)
Frequency
Gate Time (Resolution)
1 s (0.1 ppm)
0.1 s (1 ppm)
0.01 s (10 ppm)
0.001 s (100 ppm)
3 Hz-5Hz
0
0.12
0.12
0.12
5 Hz-10Hz
0
0.17
0.17
0.17
10 Hz-40Hz
0
0.20
0.20
0.20
40 Hz-100Hz
0
0.06
0.21
0.21
100 Hz-300Hz
0
0.03
0.21
0.21
300 Hz-1 kHz
0
0.01
0.07
0.07
>1kHz
0
0
0.02
0.02
[1]
Specifications are for 90 minutes warm-up and 1 s gate time.
[2] For frequency ≤300 kHz, the specification is for AC input voltage of 10% to 110% of range. For frequency >300 kHz, the specification is for AC
input voltage of 20% to 110% of range. The maximum input is limited to 750 Vrms or 8 x 10
7
Volts-Hz (whichever is less). 200 mV range is full
range input or input that is larger than the full range. For 20 mV to 200 mV inputs, multiply % of reading error by 10.
[3]
Relative to calibration standards.