User`s guide

RIGOL Chapter 2 Front Panel Operations
2-88 M300 User’s Guide
Tip
When all the three IP configuration modes are enabled, the priority of
the parameter configuration from high to low is DHCP, AutoIP” and
ManualIP.
The three IP configuration modes cannot be all disabled at the same
time.
1) DHCP (Dynamic Host Configuration Protocol)
In this mode, the DHCP server of the current network distributes network
parameters (such as the IP address) to the instrument.
Use the up/down direction keys to select DHCP and use the left/right
direction keys to enable or disable the DHCP configuration mode. By default,
DHCP is enabled.
2) AutoIP
In this mode, the instrument acquires an IP address from 169.254.0.1 to
169.254.255.254 and subnet mask 255.255.0.0 according to the current
network configuration.
Use the up/down direction keys to select AutoIP and use the left/right
direction keys to enable or disable the AutoIP configuration mode. By
default, AutoIP is disabled.
Note: to acquire an IP address using this mode, you need to disable the
DHCP mode. If DHCP is not disabled, the instrument will use the DHCP
mode automatically.
3) ManualIP
In this mode, users define the network parameters of the instrument,
including the IP address, subnet mask, default gateway and DNS (Domain
Name Service).
Use the up/down direction keys to select ManualIP and use the left/right
direction keys to enable the ManualIP configuration mode. By default,
ManualIP is disabled. At this point, you need to use the up/down direction
keys to select IP Addr, SubMask, Gateway and DNS respectively and
use the numeric keyboard to input the correct values.
Note: to use this mode to set the IP address, you need to disable the DHCP
and AutoIP modes. Otherwise, the instrument will use the DHCP or AutoIP
mode automatically.
IP Address
The format of IP address is nnn.nnn.nnn.nnn, for example, 172.16.2.13.
After enabling the ManualIP mode, use the up/down direction keys to
select IP Addr and use the numeric keyboard to set the desired IP