User`s guide

Chapter 2 Front Panel Operations RIGOL
M300 Users Guide
2-83
Module Plug
None of the current M300 modules are hot-swappable. To avoid damage, you can
select the mainframe response below. Use the up/down direction keys to select Card
Plug and use the left/right direction keys to select the way in which the instrument
responses to the hot plug
[1]
of the module.
Restart: the instrument restarts automatically. If you insert the module during a
scan, the instrument restarts and resumes the scan.
Confirm: a prompt message is displayed. At this point, users need to confirm the
module connection.
Ignore: the instrument does not execute any operation and automatically
recognize the module.
Note
[1]
: hot plug/hot swap is defined as inserting or removing the module without turning off the
instrument.
Speed Mode
Use the up/down direction keys to select Speed and use the left/right keys to select
“Normal” or “High”. By default, “Normal” is selected.
Under Highspeed mode, Singlein advanced source can be selected, but the
operation will be invalid. At this point, if initializing a scan, the instrument will enter
high speed mode and turn off the display (display the prompt message Scanning…”)
to increase the scan speed when a scan is initialized. Pressing and holding
Run/Stop can exit the scan. The keys except Run/Stop are all locked.
Brightness
Use the up/down direction keys to select Bright and use the left/right keys to adjust
the brightness between 0 and 15 (the default is 8). The current setting is stored in
non-volatile memory.