User`s guide
RIGOL Chapter 2 Front Panel Operations
2-72 M300 User’s Guide
Store and Recall
M300 allows you to save the system configuration, measurement configuration and
measurement data into the internal or external memory and recall them when
required. Press Store/Recall to enter the interface as shown in the figure below.
Figure 2-37 Store and Recall
1 Cursor
Press Cursor to move the cursor among the three areas.
File Type Area: use the left/right direction keys to select the file type to be
saved or read.
Directory Area: use the up/down direction keys to select the directory of
the file to be saved or read. C: disk is the local directory and holds 256
Mbytes storage space; D: disk is the external USB storage device.
File Area: the read, save, and delete operations are valid only when the
cursor is placed in this area.
2 Read
Press Read to read the file currently selected. This menu is valid only when the
cursor is placed in the file area and the current disk contains the specified type
of file.
Note: before reading a system configuration file, make sure that the module
currently inserted and the module used when storing the configuration file is of
the same type.
Directory Area File Type Area File Area