User`s guide

Chapter 2 Front Panel Operations RIGOL
M300 Users Guide
2-65
Alarm Channel Setting
M300 provides 4 alarm channels (Alarm1, Alarm2, Alarm3 and Alarm4). You can
assign an alarm channel to a configured channel via the channel configuration guide.
An alarm will be generated from the corresponding channel when a reading fulfills
the specified alarm condition during a scan or monitor.
When an alarm is generated, the corresponding pin of the Alarm/Ext Trig interface at
the rear panel (converted from the [RS-232/Alarms/Ext Trig] interface) outputs
a pulse with specified edge. If an alarm channel is assigned to multiple channels, the
corresponding pin outputs a pulse when any of the channel generates an alarm.
Alarm/Ext Trig Interface
Press
, , or at the front panel to set the output
mode of the corresponding pin of the Alarm/Ext Trig interface for the 4 alarm
channels when an alarm is generated.
Figure 2-32 Alarm Channel Setting Interface
1 Mode
Set the output mode of the corresponding pin when an alarm is generated. Use
the up/down direction keys to select Mode and use the left/right direction keys
to select “Latch” or “Track”.
Pin
Definition
1
Alarm 1 Output
2
Alarm 2 Output
3
Alarm 3 Output
4
Alarm 4 Output
4
2
3
1