User`s guide

RIGOL Chapter 2 Front Panel Operations
2-56 M300 User’s Guide
various operations on the channel currently selected (such as disconnect and close
the channel).
Figure 2-24 MC3324 Control Interface (Bank Joined)
If the channel currently selected is not added to the scan list and other channels of
the module are not added to the scan list, pressing OK can close the channel. At this
point, pressing OK again can open the channel. If the channel is added to the scan
list, pressing OK can close the channel and read the measurement result (before
closing this channel, the instrument opens the channel currently closed). Pressing
OK again can read the measurement result again.
1 Open
Disconnect the channel currently selected. The channel number of the channel
currently disconnected is displayed in gray, for example,
. If one or more
channels of the current module are added to the scan list, this operation is
invalid.
2 Close
Close the channel currently selected. Only one channel can be closed each time.
The channel number of the channel currently closed is displayed in green, for
example,
. If one or more channels of the current module are added to the
scan list, this operation is invalid.
3 Read
Read the measurement value of the current channel. This operation is only
applicable to channels added to the scan list.
4 Module Reset
Initialize the status of each channel of the current module. If one or more
channels of the current module are added to the scan list, this operation is