Specifications

Chapter 2 Command System RIGOL
M300 Programming Guide 2-79
MEASure:VOLTage:AC?
Syntax MEASure:VOLTage:AC?
[{<range>|AUTO|MIN|MAX|DEF}[,{<resolution>|MIN|MAX|DEF}],] (@<scan_list>)
Description Configure the specified channels to the ACV measurement function with the specified range
and resolution, start a scan and read the readings.
Parameters
Name
Type
Range
Default
<range> Numeric
Any numeric value between MIN and MAX. The
final range is decided by the
Principle of setting
with greater value
”.
For MC3232 and MC3264 module, the standard
values of the range: {200mv|2V|20V|150V}.
wherein, MIN=200mV, MAX=150V, DEF=AUTO.
For MC3120, MC3132, MC3164 and MC3324
module, the standard values of the range:
{200mV|2V|20V|200V|300V}, wherein,
MIN=200mV, MAX=300V, DEF=AUTO.
AUTO
<resolution> Numeric
Can receive any numeric value, but the resolution is fixed at
6
1
/
2
digits.
<scan_list>
Scan
List
One or more channels (only for the multiplexer
channels), the rules are as follows:
(@101): channel 01 on the module in Slot1;
(@101:103): channel 01 through 03 on the module
in Slot1;
(@101:103,301): channel 01 through 03 on the
module in Slot1 and channel 01 on the module in
Slot3.
None
Explanation This command is equivalent to the CONFigure:VOLTage:AC command followed by a
READ? command.
You can select autoranging to allow the instrument to automatically select a proper
measurement range or you can select a fixed range to set the range manually.
Autoranging rule: for signals under test that is between 10%*Range and
110%*Range, the instrument automatically selects Range as the current range.
When <range> is set to DEF or AUTO, an error will be generated if <resolution> is set
to a numeric value, because the instrument cannot calculate the integration time
accurately (especially when the input signal is continuously changing) when the
autoranging is combined with a numeric resolution. If your application requires
autoranging, be sure to specify "DEF" for <resolution> or omit the parameter.
If the input signal is greater than can be measured on the selected range, the
instrument gives an overload indication: "OVERLOAD" from the front panel or
"±9.9E+37" from the remote interface.
<scan_list> overwrites the current scan list.
Return
Format
The query returns the readings in scientific notation. Multiple return values are separated
by commas.
Example MEAS:VOLT:AC? AUTO,DEF,(@101)
The query returns +9.689453687E-02.
Related
commands
CONFigure:VOLTage:AC
READ?
ROUTe:CHANnel:DELay
[SENSe:]VOLTage:AC:BANDwidth