MEASure:TEMPerature?
Syntax MEASure:TEMPerature?
{<probe_type>|DEF},{<type>|DEF}[,1[,{<resolution>|MIN|MAX|DEF}]],(@<scan_list>)
Description Configure the specified channels to the temperature measurement function, start a scan
and read the readings.
Parameters
Name
Type
Range
Default
<probe_type>
Discrete
{TCouple|THERmistor|RTD|FRTD}
TC
<type> Discrete
TCouple
{B|E|J|K|N|R|S|T}
J
THERmistor
{2252|3000|5000|10000|30000}
5000
RTD
{85|89|91|92}
85
FRTD
{85|89|91|92}
85
<resolution> Numeric
Can receive any numeric value. This parameter is unessary
for the command. The instrument only receives the
parameter but does not respond to it.
<scan_list>
Scan
List
One or more channels (only for the multiplexer
channels), the rules are as follows:
(@101): channel 01 on the module in Slot1;
(@101:103): channel 01 through 03 on the
module in Slot1;
(@101:103,301): channel 01 through 03 on the
module in Slot1 and channel 01 on the module in
Slot3.
None
Explanation This command is equivalent to the CONFigure:TEMPerature command followed by the
READ? command.
For all the channels of MC3164 and MC3264, <probe_type> can not be set to FRTD
(4-wire RTD).
If <probe_type> is specified as FRTD, channel n is paired with channel n+16
(MC3132 and MC3232, wherein the range of n is from 1 to 16) or n+10 (MC3324 and
MC3120, wherein the range of n is from 1 to 10) automatically. Channel n is used to
connect the source terminal of the DMM and channel n+16 or channel n+10 is used to
connect the sense terminal of the DMM. The paired channels cannot be configured.
If the input signal is greater than can be measured on the selected range, the
instrument gives an overload indication: "OVERLOAD" from the front panel or
"±9.9E+37" from the remote interface.
<scan_list> overwrites the current scan list.
Return
Format
The query returns the readings in scientific notation. Multiple return values are separated
by commas.
Example MEAS:TEMP? DEF,DEF,1,DEF,(@101)
The query returns +2.411291906E+01.
Related
commands
CONFigure:TEMPerature