Specifications
RIGOL Chapter 2 Command System
2-60 M300 Programming Guide
*SAV
*RCL
Syntax *SAV {0|1|2|3|4|5}
*RCL {0|1|2|3|4|5}
Description The *SAV command stores the current measurement configuration to the specified
location and overwites any file previously stored in the current storage location (no error
is generated). The *RCL command recalls the measurement file stored in the specified
location and overwrites the current measurement configuration.
Explanation The instrument has six storage locations in the non-volatile memory (refer to
Appendix C: Non-volatile Memory) for the measurement configuration. You can store
the current measurement configuration into the measurement configuration file in
any of the six storage locations (named location 0, 1, 2, 3, 4, and 5. The default file
names corresponding to the locations are 0_STATE0, 1_STATE1, 2_STATE2,
3_STATE3, 4_STATE4 and 5_STATE5).
The extension of the measurement configuration file is “.mfg”.
The measurement configuration stores the scan list settings (include the trigger
count, trigger mode, interval for auto trigger mode and trigger edge type for external
trigger) and the channel configurations (include the measurement configuration,
scalling configuration, alarm configuration and advanced configuration).
When shipped from the factory, all the measurement configuration files in the six
storage locations are empty.
If the measurement configuration file of the specified location dose not exist
(Sending the
SYSTem:SECurity[:IMMediate] command will delete all the files of the
memory) , you can send the
*SAV command to create the measurement
configuration file of the specified location.
A Factory Reset (the
*RST command) does not affect the measurement configuration
file stored in the specified location. Sending the
SYSTem:SECurity[:IMMediate]
command will delete all the measurement configuration files of the memory.
Related
Commands
MEMory:STATe:DELete
MEMory:STATe:NAME
MEMory:STATe:VALid?