Specifications
Chapter 2 Command System RIGOL
M300 Programming Guide 2-53
FORMat:READing:TIME
Syntax FORMat:READing:TIME <mode>
FORMat:READing:TIME?
Description Disable or enable the time stamp in the return redings of the scan result query commands
(such as the
READ? command, the R? command and the FETCh? command).
Parameters
Name
Type
Range
Default
<mode>
Bool
{OFF|0|ON|1}
OFF
Explanation The setting applies to all the readings aquired in the scan. You cannot set the reading
format on a per-channel basis.
This command operates in conjunction with the other
FORMat Command Subsystem
commands (they are not mutually exclusive).
If enabled, the time stamp information is shown either in absolute time (time of day
with date) or relative time (time in seconds relative to the start of the scan) as set by
the
FORMat:READing:TIME:TYPE command.
This setting is stored in the volatile memory (refer to
Appendix D: Volatile memory)
and will be disabled (OFF) when the power is turned off or after a Factory Reset (the
*RST command).
Return
Format
The query returns 0 (OFF) or 1 (ON).
Example
FORM:READ:TIME ON
FORM:READ:TIME?
The query returns 1.
Related
commands
FORMat Command Subsystem
R?
READ?
FETCh?