Specifications
Chapter 2 Command System RIGOL
M300 Programming Guide 2-43
DIAGnostic Command Subsystem
DIAGnostic:DMM:CYCLes?
DIAGnostic:DMM:CYCLes:CLEar
DIAGnostic:PEEK:SLOT:DATA
DIAGnostic:POKE:SLOT:DATA
DIAGnostic:RELay:CYCLes?
DIAGnostic:RELay:CYCLes:CLEar
DIAGnostic:DMM:CYCLes?
Syntax
DIAGnostic:DMM:CYCLes?
Description Query the cycle count of the 6 relays on the DMM module. The query returns six numbers
indicating the cycle count on relays 1, 2, 3, 4, 5 and 6 (which correspond to relays K1, K2,
K3, K4, K5 and K6 respectively).
Explanation These relays open or close when the function or range of the DMM module is
changed.
The instrument generates an error if the DMM module is not installed when sending
this command.
Return
Format
The query returns 6 numbers, indicating the cycle counts of the 6 relays respectively.
Example
DIAG:DMM:CYCL?
The query returns +0,+0,+0,+0,+37,+0.
DIAGnostic:DMM:CYCLes:CLEar
Syntax
DIAGnostic:DMM:CYCLes:CLEar {1|2|3|4|5|6}
Description
Reset the cycle counts of the 6 relays on the DMM module.
Parameters
Name
Type
Range
Default
{1|2|3|4|5|6}
Discrete
1|2|3|4|5|6
None
Related
command
DIAGnostic:DMM:CYCLes?