Specifications

Chapter 2 Command System RIGOL
M300 Programming Guide 2-37
CONFigure:VOLTage[:DC]
Syntax CONFigure:VOLTage[:DC]
[{<range>|AUTO|MIN|MAX|DEF}[,{<resolution>|MIN|MAX|DEF}],](@<scan_list>)
Description Configure the specified channels to the DCV measurement function with the specified
range and resolution. This command does not initiate the scan.
Parameters
Name
Type
Range
Default
<range> Numeric
Any numeric value between MIN and MAX. The
final range is decided by the
Principle of setting
with greater value
”.
For the MC3232 and MC3264 modules, the
standard values of the range:
{200mv|2V|20V|150V}.
Wherein, MIN=200mV, MAX=150V, DEF=AUTO.
For the MC3120, MC3132, MC3164 and MC3324
modules, the standard values of the range:
{200mV|2V|20V|200V|300V}. Wherein,
MIN=200mV, MAX=300V, DEF=AUTO.
AUTO
<resolution> Numeric
Can receive any numeric value between
0.03ppm×<range> and 3ppm×<range>. The
final resolution is decided by the
Principle of
setting with smaller value
”.
The standard values of the resolution: refer to the
Explanation.
0.3ppm×
<range>
<scan_list>
Scan
List
One or more channels (only for the multiplexer
channels), the rules are as follows:
(@101): channel 01 on the module in Slot1;
(@101:103): channel 01 through 03 on the
module in Slot1;
(@101:103,301): channel 01 through 03 on the
module in Slot1 and channel 01 on the module in
Slot3.
None
Explanation <range> is used to set the range. You can select autoranging to allow the instrument
to automatically select a proper measurement range or you can select a fixed range to
set the range manually.
AAutoranging rule: for signals under test that is between 10%*Range and
110%*Range, the instrument automatically selects Range as the current range.
When <range> is set to DEF or AUTO, an error will be generated if <resolution> is set
to a numeric value, because the instrument cannot calculate the integration time
accurately (especially when the input signal is continuously changing) when the
autoranging is combined with a numeric resolution. If your application requires
autoranging, be sure to specify "DEF" for <resolution> or omit the parameter.
<resolution> is related to the current integration time and range (<range>). The
relations are as shown in the table below.
Integration Time
Resolution
0.02PLC
3ppm× <range> (MAX)
0.2PLC
0.7ppm× <range>
1PLC
0.3ppm× <range> (DEF)
2PLC
0.2ppm× <range>
10PLC
0.1ppm× <range>
20PLC
0.06ppm× <range>