Specifications

RIGOL Chapter 2 Command System
2-32 M300 Programming Guide
<resolution> is related to the current integration time and range (<range>). The
relations are as shown in the table below.
Integration Time
Resolution
0.02PLC
3ppm× <range> (MAX)
0.2PLC
0.7ppm× <range>
1PLC
0.3ppm× <range> (DEF)
2PLC
0.2ppm× <range>
10PLC
0.1ppm× <range>
20PLC
0.06ppm× <range>
100PLC
0.035ppm× <range>
200PLC
0.03ppm× <range> (MIN)
If the input signal is greater than can be measured on the selected range, the
instrument gives an overload indication: "OVERLOAD" from the front panel or
"±9.9E+37" from the remote interface.
The
*RST command will clear the scan list and set all the measurement parameters to
their factory settings. The Instrument Preset (the
SYSTem:PRESet command) will not
clear the scan list; however, this command will clear the data in the reading memory.
<scan_list> overwrites the current scan list.
Example
CONFigure:RES AUTO,DEF,(@101)
Related
commands
CONFigure?
INITiate
READ?