Specifications
Chapter 2 Command System RIGOL
M300 Programming Guide 2-31
CONFigure:FRESistance
CONFigure:RESistance
Syntax CONFigure:FRESistance
[{<range>|AUTO|MIN|MAX|DEF}[,{<resolution>|MIN|MAX|DEF}],](@<scan_list>)
CONFigure:RESistance
[{<range>|AUTO|MIN|MAX|DEF}[,{<resolution>|MIN|MAX|DEF}],](@<scan_list>)
Description Configure the specified channels to the 2-wire resistance or 4-wire resistance
measurement function with the specified range and resolution, but do not initiate the
scan.
Parameters
Name
Type
Range
Default
<range> Numeric
Any numeric value between MIN and MAX.
The final range is decided by the “
Principle of
setting with greater value
”.
The standard values of the range:
{200Ω|2kΩ|20kΩ|200kΩ|1MΩ|10MΩ|100MΩ}
Wherein, MIN=200Ω, MAX=100MΩ,
DEF=AUTO.
AUTO
<resolution> Numeric
Any numeric value between
0.03ppm×<range> and 3ppm×<range>. The
final resolution is decided by the “
Principle of
setting with smaller value
”.
The standard values of the resolution: refer to
the “Explanation”.
0.3ppm×
<range>
<scan_list>
Scan
List
One or more channels (only for the
multiplexer channels), the rules are as
follows:
(@101) :channel 01 on the module in Slot1;
(@101:103): channel 01 through 03 on the
module in Slot1;
(@101:103,301): channel 01 through 03 on
the module in Slot1 and channel 01 on the
module in Slot3;
None
Explanation For MC3164 and MC3264, the 4-wire resistance measurement function is not
supported.
For the 4-wire resistance measurement function, channel n is paired with channel
n+16 (MC3132 and MC3232, wherein the range of n is from 1 to 16) or n+10
(MC3324 and MC3120, wherein the range of n is from 1 to 10) automatically. Channel
n is used to connect the source terminal of the DMM and channel n+16 or channel
n+10 is used to connect the sense terminal of the DMM. The paired channels cannot
be configured.
<range> is used to set the range. You can select autoranging to allow the instrument
to automatically select a proper measurement range or you can select a fixed range
to set the range manually.
Autoranging rule: for signals under test that is between 10%*Range and
110%*Range, the instrument automatically selects Range as the current range.
When <range> is set to DEF or AUTO, an error will be generated if <resolution> is
set to a numeric value, because the instrument cannot calculate the integration time
accurately (especially when the input signal is continuously changing) when the
autoranging is combined with a numeric resolution. If your application requires
autoranging, be sure to specify "DEF" for <resolution> or omit the parameter.