Specifications

RIGOL Chapter 2 Command System
2-30 M300 Programming Guide
CONFigure:FREQuency
CONFigure:PERiod
Syntax CONFigure:FREQuency
[{<range>|AUTO|MIN|MAX|DEF}[,{<resolution>|MIN|MAX|DEF}],](@<scan_list>)
CONFigure:PERiod
[{<range>|AUTO|MIN|MAX|DEF}[,{<resolution>|MIN|MAX|DEF}],](@<scan_list>)
Description Configure the specified channels to the frequency or period measurement function with the
specified gate time, but do not initiate the scan.
Parameters
Name
Type
Range
Default
<range> Numeric
Any numeric value. This parameter is unnecessary
for the frequency and period measurements, the
instrument only receives the parameter but does
not respond to it.
AUTO
<resolution> Numeric
Any numeric value between MIN and MAX. The final
gate time is decided by the
Principle of setting
with greater value
”.
The standard values of the gate time:
{1ms|10ms|100ms|1s}
Wherein, MIN=1ms, MAX=1s, DEF=100ms.
100ms
<scan_list>
Scan
List
One or more channels (only for the multiplexer
channels), the rules are as follows:
(@101): channel 01 on the module in Slot1;
(@101:103): channel 01 to channel 03 on the
module in Slot1;
(@101:103,301): channel 01 to channel 03 on the
module in Slot1 and channel 01 on the module in
Slot3;
None
Explanation If the input signal is greater than can be measured on the selected range, the
instrument gives an overload indication: "OVERLOAD" from the front panel or
"±9.9E+37" from the remote interface.
The
*RST command will clear the scan list and set all the measurement parameters to
their factory settings. The Instrument Preset (the
SYSTem:PRESet command) will not
clear the scan list; however, this command will clear the data in the reading memory.
<scan_list> overwrites the current scan list.
Example
CONF:FREQ MAX,DEF,(@101)