Specifications
RIGOL Chapter 3 Application Examples
3-2 M300 Programming Guide
Scan List Configuration
To Configure the Channels
Conditions:
MC3324 module in Slot1
MC3132 module in Slot2
MC3534 module in Slot3
MC3065 module in Slot4
MC3648 module in Slot5
Requirements
Use the SCPI commands to realize the following functions:
Configure the channels as follows and add these channels to the scan list to measure a variety of input
signals.
Channel
Configuration
Measurement Function
Parameter
101
Temperature, RTD, 85
Integration time: 10PLC; R0=50Ω;
102
VOLTage:DC
Range: AUTO; Integration time: 100PLC; Enable the input
resistance mode;
Scaling: A=1.001, B=1.1, C=0.01, x1=0.05;
Alarm setting: HI=10V, LO=0.7V, alarm channel (Alarm1);
103:104
VOLTage:AC
Range: 20V; AC Filter: 3Hz;
Alarm setting: HI=5V, alarm channel (Alarm2);
105
Resistance
Range: 100MΩ; Enable the offset compensation function;
Scaling: A=1.001, B=1.1, C=0.01, x1=10, unit = k;
Alarm setting: HI=90MΩ, alarm channel (Alarm3);
116
Temperature, FRTD, 92
Integration time: 200PLC;
Enable the offset compensation function; R0=50Ω;
107:108
FResistance
Range:AUTO;
Integration time: 0.2PLC; Enable Auto Zero;
109:110
Frequency
Voltage range: AUTO; Gate time: 10ms;
Alarm setting: HI=3kHz, alarm channel (Alarm4);
111:112
Period
Voltage range: 300V; AC Filter: 200Hz;
113
Temperature, TC, J
Integration time: 0.5ms; unit: F
Scaling: A=1.001, B=1.1, C=0.01, x1=0.01;
Alarm setting: HI=30F, alarm channel (Alarm1);
Reference source: external; Enable Auto Zero; Enable T/C
Check;
114
Temperature, THER, 5000
Integration time: 2PLC; Enable Auto Zero;
Alarm setting: LO=10°C, alarm channel (Alarm2) ;
115
Temperature,RTD,91
Integration time: 20PLC; R0=50Ω;
119
Temperature,TC,K
Alarm setting: LO=10°C, alarm channel (Alarm3) ;
Reference source: internal;
120
Temperature,TC,S
Reference source: fixed;
121:122
CURRent:DC
Range: AUTO; Integration time: 1ms;
Alarm setting: HI=0.6A, alarm channel (Alarm4) ;
123
CURRent:AC
Range: 1A; AC Filter: 3Hz; Channel delay: 1ms;
124
Anysensor:CURRent
201
Anysensor:VOLTage
Integration time: 3ms; Unit: mV
Scaling: A=1.001, B=1.1, C=0.01, x1=0.03;
202
Anysensor:Resistance
Enable Auto Zero;
203
Anysensor:FResistance
Alarm setting: LO=1kΩ, alarm channel (Alarm1); Enable
the offset compensation function;
204
Anysensor:Frequency