Specifications

RIGOL Chapter 2 Command System
2-230 M300 Programming Guide
The trigger mode is set to IMMediate (auto) automatically after a Factory Reset (send
the
*RST command). An Instrument Preset (send the SYSTem:PRESet command) or
Card Reset (send the SYSTem:CPON command) does not affect the setting.
Return
Format
The query returns BUS, IMM, EXT, ALAR1, ALAR2, ALAR3, ALAR4 or ABS.
Example
TRIG:SOUR ABS
TRIG:SOUR?
The query returns ABS.
Related
commands
TRIGger:ABSolute
TRIGger:TIMer
TRIGger:TIMer
Syntax TRIGger:TIMer {<seconds>|MIN|MAX}
TRIGger:TIMer? [{MIN|MAX}]
Description
Set the scan interval in the TIMer trigger mode.
Parameters
Name
Type
Range
Default
<seconds>
Numeric
0 to 359999.999s
0
Explanation The time interval defines the time from the start of a scan to the start of the next scan
in the auto scan mode, as shown below.
If the scan interval specified is shorter than the time required to perform a complete
scan of the scan list, the instrument will scan continuously as fast as possible (no error
is generated).
Send the TRIG:TIM? MIN command to query the minimum interval and the query
returns +0.000000000E+00.
Send the TRIG:TIM? MAX command to query the maximum interval and the query
returns +3.599999990E+05.
The
CONFigure Command Subsystem and MEASure Command Subsystem commands
automatically set the scan interval to 0.
The instrument sets the scan interval to 0 after a Factory Reset (send the
*RST
command). An Instrument Preset (send the
SYSTem:PRESet command) or Card Reset
(send the SYSTem:CPON command) does not affect the setting.
Return
Format
The query returns the current scan interval in scientific notation, for example,
+3.600000000E+04. The unit is s.
Example TRIG:TIM 36000
TRIG:TIM?
The query returns +3.600000000E+04.
Related
command
TRIGger:SOURce
……
Interval
A scan