SYSTem:REMote
Syntax
SYSTem:REMote
Description
Place the instrument in the remote mode.
Explanation
In the remote mode, all the keys (except
, and ) on the front panel are
disabled.
Example
SYST:REM
Related
commands
SYSTem:LOCal
SYSTem:RWLock
SYSTem:RWLock
Syntax
SYSTem:RWLock
Description
Place the instrument in the remote locked mode.
Explanation
In the locked mode, all the keys (except
) on the front panel are disabled. You can
press and hold or send the SYSTem:LOCal command to unlock the instrument.
Example
SYST:RWL
Related
commands
SYSTem:LOCal
SYSTem:REMote
SYSTem:SECurity[:IMMediate]
Syntax
SYSTem:SECurity[:IMMediate]
Description Clear all the data (except the MAC address, calibration parameters and serial number) in
the memory. This command usually restore the instrument to the initial state.
Explanation This command restore the instrument to the factory settings (refer to the *RST
command).
All the I/O parameters (such as the IP address) are returned to their factory settings.
This command will not clear an attached USB storage device.
The command clears and sanitizes all user files on the internal file system.
Example
SYST:SEC
Related
command
*RST
SYSTem:SERIal?
Syntax
SYSTem:SERIal?
Description
Query the serial number of M300.
Return
Format
The query returns the serial number in the form of "M300XXXXXXXXX".
Example
SYSTem:SERIal?
The query returns M300123123123.
Related
command
SYSTem Command Subsystem