Specifications

RIGOL Chapter 2 Command System
2-212 M300 Programming Guide
SYSTem:CPON
Syntax
SYSTem:CPON <slot>
Description
Reset the module in the specified slot.
Parameters
Name
Type
Range
Default
<slot>
Discrete
{100|200|300|400|500|ALL}
None
Explanation The effect of this command on different module is follows.
Module
Effect
MC3120
MC3132
MC3232
MC3164
MC3264
MC3324
If any channel is configured into the scan list, this command
has no effect. If no channel is configured into the scan lsit, this
command opens all the channels.
MC3416
This command opens all the channels.
MC3534
If any channel is configured into the scan lsit, this command
has no effect.
If no channel is configured into the scan list, the DIO channels
will be congifured as 8-bit digital input ports; the count of each
TOT channel will be cleard, its reading mode will be set to
READ and the output voltage of each DAC channel will be
congfigured to 0V.
MC3648
This command opens all the channels.
This command does not reset the DMM module.
If the instrument is scanning or any channel of the specified module is added into the
scan list, an error will be generated when sending this command.
To reset the modules in all the five slots, set <slot> to ALL.
Example
SYST:CPON 200
Related
commands
*RST
SYSTem:PRESet