Specifications

Chapter 2 Command System RIGOL
M300 Programming Guide 2-209
SYSTem:COMMunicate:LAN:UPDate
Syntax
SYSTem:COMMunicate:LAN:UPDate
Description Update the LAN parameters. Disconnect all the LAN and Web connections and restart the
LAN interface with the current LAN parameters.
Explanation Executing this command updates the settings of the following commands.
SYSTem:COMMunicate:LAN:AUTOip
SYSTem:COMMunicate:LAN:DHCP
SYSTem:COMMunicate:LAN:MANUip
SYSTem:COMMunicate:LAN:DNS
SYSTem:COMMunicate:LAN:GATEway
SYSTem:COMMunicate:LAN:IPADdress
SYSTem:COMMunicate:LAN:SMASk
When the LAN parameters set are invalid, the LAN cannot be connected normally
when this command is sent.
Related
commands
SYSTem:COMMunicate:LAN:DHCP
SYSTem:COMMunicate:LAN:AUTOip
SYSTem:COMMunicate:LAN:MANUip
SYSTem:COMMunicate:LAN:DNS
SYSTem:COMMunicate:LAN:GATEway
SYSTem:COMMunicate:LAN:IPADdress
SYSTem:COMMunicate:LAN:SMASk
SYSTem:COMMunicate:RS232:BAUD
Syntax SYSTem:COMMunicate:RS232:BAUD {4800|9600|19200|38400|57600|115200}
SYSTem:COMMunicate:RS232:BAUD?
Description
Set or query the baud rate of the RS232 interface and the unit is Baud.
Parameters
Name Type Range Default
{4800|9600|19200|
38400|57600|115200}
Discrete
4800|9600|19200|
38400|57600|115200
None
Explanation The baud rate of the RS232 interface is stored in the non-volatile memory (refer to
Appendix C: Non-volatile Memory) and does not change when the power has been off, after
a Factory Reset (send the
*RST command) or after an Instrument Preset (send the
SYSTem:PRESet command).
Return
Format
The query returns the current baud rate, for example, 19200.
Example
SYST:COMM:RS232:BAUD 19200
SYST:COMM:RS232:BAUD?
The query returns 19200.
Related
commands
SYSTem:COMMunicate:RS232:FLOWcontrol
SYSTem:COMMunicate:RS232:PARIty
SYSTem:COMMunicate:RS232:PRINt:STATe