Specifications
RIGOL Chapter 2 Command System
2-180 M300 Programming Guide
[SENSe:]VOLTage:AC:RESolution
Syntax [SENSe:]VOLTage:AC:RESolution {<resolution>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]VOLTage:AC:RESolution? [{(@<ch_list>)|MIN|MAX}]
Description
Set or query the resolution of the ACV measurements on the specified channels.
Parameters
Name
Type
Range
Default
<resolution> Numeric
Can receive any numeric value,
but the resolution is fixed at 6
1
/
2
digits.
<ch_list>
Channel
List
One or more channels (only for the
multiplexer channels), the rules are as
follows:
(@101): channel 01 on the module in
Slot1;
(@101:103): channel 01 through 03 on
the module in Slot1;
(@101:103,301): channel 01 through 03
on the module in Slot1 and channel 01 on
the module in Slot3.
If the parameter
is omitted, this
command will be
applied to the
whole scan list.
Explanation When the range is set to autoranging, an error will be generated when <resolution> is set
to a numeric value (except MIN and MAX).
Return
Format
The query returns the resolution in the form of scientific notation for each channel
specified. Multiple responses are separated by commas.
Related
commands
[SENSe:]VOLTage:AC:RANGe
[SENSe:]VOLTage:AC:RANGe:AUTO
CONFigure:VOLTage:AC
MEASure:VOLTage:AC?