Specifications
RIGOL Chapter 2 Command System
2-166 M300 Programming Guide
[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE
Syntax [SENSe:]TEMPerature:TRANsducer:TCouple:TYPE <type>[,(@<ch_list>)]
[SENSe:]TEMPerature:TRANsducer:TCouple:TYPE? [(@<ch_list>)]
Description Set or query the thermocouple type of the thermocouple temperature measurements on
the specified channels.
Parameters
Name
Type
Range
Default
<type>
Discrete
{B|E|J|K|N|R|S|T}
J
<ch_list>
Channel
List
One or more channels (only for the
multiplexer channels), the rules are as
follows:
(@101): channel 01 on the module in
Slot1;
(@101:103): channel 01 through 03 on
the module in Slot1;
(@101:103,301): channel 01 through 03
on the module in Slot1 and channel 01 on
the module in Slot3.
If the parameter
is omitted, this
command will be
applied to the
whole scan list.
Explanation The instrument sets the thermocouple type to "J" after a Factory Reset (send the *RST
command). An Instrument Preset (send the
SYSTem:PRESet command) or Card Reset
(send the SYSTem:CPON command) does not affect the current thermalcouple type.
Return
Format
The query returns the thermocouple type for each channel specified. Multiple return values
are separated by commas.
Example TEMP:TRAN:TC:TYPE K,(@101:103)
TEMP:TRAN:TC:TYPE? (@101:103,205)
The query returns K,K,K,J.
Related
command
SENSe Command Subsystem