[SENSe:]FREQuency:VOLTage:RANGe
[SENSe:]PERiod:VOLTage:RANGe
Syntax [SENSe:]FREQuency:VOLTage:RANGe {<range>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]FREQuency:VOLTage:RANGe? [{(@<ch_list>)|MIN|MAX}]
[SENSe:]PERiod:VOLTage:RANGe {<range>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]PERiod:VOLTage:RANGe? [{(@<ch_list>)|MIN|MAX}]
Description Set or query the voltage ranges of the frequency or period measurements on the specified
channels.
Parameters
Name
Type
Range
Default
<range> Discrete
Any numeric value between MIN and MAX.
The final gate time is decided by the
Principle of setting with greater value”.
For MC3232 and MC3264, the standard
values of the range are
{200mV|2V|20V|150V};
wherein, MIN=200mV, MAX=150V.
For MC3120, MC3132, MC3164 and MC3324,
the standard values of the range are
{200mV|2V|20V|200V|300V|MIN|MAX};
wherein, MIN=200mV, MAX=300V.
Auto
<ch_list>
Channel
List
One or more channels (only for the
multiplexer channels), the rules are as
follows:
(@101): channel 01 on the module in Slot1;
(@101:103): channel 01 through 03 on the
module in Slot1;
(@101:103,301): channel 01 through 03 on
the module in Slot1 and channel 01 on the
module in Slot3.
If the parameter
is omitted, this
command will be
applied to the
whole scan list.
Explanation Selecting a specific range will disable the autoranging on the specified channels.
The
CONFigure:FREQuency (CONFigure:PERiod) or MEASure:FREQuency?
(
MEASure:PERiod?) command automatically enables the autoranging if the first
parameter is AUTO, DEF or omitted.
The instrument selects autoranging after a Factory Reset (send the
*RST command).
An Instrument Preset (send the
SYSTem:PRESet command) or Card Reset (send the
SYSTem:CPON command) does not affect the current range setting.
Return
Format
The query returns the voltage range in scientific notation for each channel specified.
Multiple return values are separated by commas.
Example FREQ:VOLT:RANG 20,(@301)
FREQ:VOLT:RANG? (@301)
The query returns +2.00000000E+01.
You can replace FREQ with PER to set or query the voltage ranges of the period
measurements on the specified channels.
Related
commands
SENSe Command Subsystem
[SENSe:]FREQuency:VOLTage:RANGe:AUTO
[SENSe:]PERiod:VOLTage:RANGe:AUTO