Specifications

RIGOL Chapter 2 Command System
2-134 M300 Programming Guide
[SENSe:]CURRent:AC:RESolution
Syntax [SENSe:]CURRent:AC:RESolution {<resolution>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]CURRent:AC:RESolution? [{(@<ch_list>)|MIN|MAX}]
Description
Set or query the resolution of the ACI measurements on the specified channels.
Parameters
Name
Type
Range
Default
<resolution> Numeric
Can receive any numeric value,
but the resolution is fixed at 6
1
/
2
digits.
<ch_list>
Channel
List
One or more channels (only for channel 21
to channel 24 of MC3324), the rules are as
follows:
(@121): channel 21 on the module in
Slot1;
(@121:123): channel 21 through 23 on the
module in Slot1;
(@121:123,324): channel 21 through 23
on the module in Slot1 and channel 24 on
the module in Slot3.
If the
parameter is
omitted, this
command will
be applied to
the whole scan
list.
Explanation When the range is set to autoranging, an error will be generated when <resolution> is set
to a numeric value (except MIN and MAX).
Return
Format
The query returns the resolution of the specified channel in scientific notation. Multiple
return values are separated by commas.
Related
commands
[SENSe:]CURRent:AC:RANGe
[SENSe:]CURRent:AC:RANGe:AUTO
CONFigure:CURRent:AC
MEASure:CURR:AC?