Specifications
Chapter 2 Command System RIGOL
M300 Programming Guide 2-119
[SENSe:]ANYSensor:FRESistance:APERture
[SENSe:]ANYSensor:RESistance:APERture
Syntax [SENSe:]ANYSensor:FRESistance:APERture {<time>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]ANYSensor:FRESistance:APERture? [{(@<ch_list>)|MIN|MAX}]
[SENSe:]ANYSensor:RESistance:APERture {<time>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]ANYSensor:RESistance:APERture? [{(@<ch_list>)|MIN|MAX}]
Description Set the integration time via the aperture time mode for the 2-wire or 4-wire resistance
measurements (the anysensor measurement function) on the specified channels.
Parameters
Name
Type
Range
Default
<time> Numeric
Any numeric value between MIN and MAX.
Wherein: MIN=33 µs, MAX=4 s.
None
<ch_list>
Channel
List
One or more channels (only for the
multiplexer channels), the rules are as
follows:
(@101): channel 01 on the module in Slot1;
(@101:103): channel 01 through 03 on the
module in Slot1;
(@101:103,301): channel 01 through 03 on
the module in Slot1 and channel 01 on the
module in Slot3.
If the parameter
is omitted, this
command will be
applied to the
whole scan list.
Explanation For the 4-wire resistance measurement function, channel n is paired with channel
n+16 (MC3132 and MC3232, wherein the range of n is from 1 to 16) or n+10 (MC3324
and MC3120, wherein the range of n is from 1 to 10) automatically. Channel n is used
to connect the source terminal of the DMM and channel n+16 or channel n+10 is used
to connect the sense terminal of the DMM. The paired channels cannot be configured.
The
CONFigure:ANYSensor, MEASure:ANYSensor? or
[SENSe:]ANYSensor:FRESistance:NPLC command automatically disables the aperture
time mode and enables the power line cycles mode.
You can use MIN or MAX to set <time>. When MIN is selected, the instrument selects
the maximum resolution; when MAX is selected, the instrument selects the minimum
resolution.
The aperture time mode is disabled after a Factory Reset (send the
*RST command).
An Instrument Preset (send the
SYSTem:PRESet command) or Card Reset (send the
SYSTem:CPON command) does not affect the current integration time parameter.
Return
Format
The query returns the integration time in scientific notation for each channel specified.
Multiple return values are separated by commas.
Example
ANYS:FRES:APER 1,(@201,202)
ANYS:FRES:APER? (@201,202)
The query returns +1.00000000E+00,+1.00000000E+00.
You can replace FRES with RES to set the integration time via the aperture time mode for
the 2-wire resistance measurements.
Related
commands
SENSe Command Subsystem
CONFigure:ANYSensor
MEASure:ANYSensor?
[SENSe:]ANYSensor:FRESistance:NPLC
[SENSe:]ANYSensor:RESistance:NPLC