Specifications

RIGOL Chapter 2 Command System
2-118 M300 Programming Guide
[SENSe:]ANYSensor:FREQuency:RANGe:LOWer
Syntax [SENSe:]ANYSensor:FREQuency:RANGe:LOWer {<frequency>|MIN|MAX}[,(@<ch_list>)]
[SENSe:]ANYSensor:FREQuency:RANGe:LOWer? [{(@<ch_list>)|MIN|MAX}]
Description Set the AC filter parameter for the frequency measurements (anysensor function) of the
specified channels.
Parameters
Name
Type
Range
Default
<frequency> Numeric
Any numeric value between MIN and
1000000. The final AC filter parameter is
decided by the
Principle of setting with
samller value
”.
The standard values of the range:
{3|20|200}
Wherein: MIN=3Hz, MAX=200Hz.
20Hz
<ch_list>
Channel
List
One or more channels (only for the
multiplexer channels), the rules are as
follows:
(@101): channel 01 on the module in
Slot1;
(@101:103): channel 01 through 03 on the
module in Slot1;
(@101:103,301): channel 01 through 03
on the module in Slot1 and channel 01 on
the module in Slot3.
If the
parameter is
omitted, this
command will
be applied to
the whole scan
list.
Explanation The CONFigure:ANYSensor or MEASure:ANYSensor? command automatically select
the 20 Hz filter.
During the measurement, the filter type is determined by the frequency of the input
signal of the current channel as shown in the table below.
Input Frequency
AC Filter Type
3 Hz to 300 kHz
3 Hz (slow)
20 Hz to 300 kHz
20 Hz (medium)
200 Hz to 300 kHz
200 Hz (fast)
The instrument selects the 20 Hz filter automatically after a Factory Reset (send the
*RST command). An Instrument Preset (send the SYSTem:PRESet command) or Card
Reset (send the
SYSTem:CPON command) does not affect the current AC filter
parameter.
Return
Format
The query returns the AC filter parameter in scientific notation. Multiple return values are
separated by commas.
Example ANYS:FREQ:RANG:LOW 200,(@301)
ANYS:FREQ:RANG:LOW? (@301)
The query returns 2.000000000E+02.
Related
commands
SENSe Command Subsystem
CONFigure:ANYSensor
MEASure:ANYSensor?